Dear fellow spinlanders,
I am trying to keep this Varian Unity/Inova 500 from 1997 alive, and have had
various issues along the way. The spectrometer doesn't get much use,
and at some point will need to be retired. As of late, I've been
presented with a new issue.
The day I came back from vacation, I believe things on the spect looked normal,
however just before lunch I had noticed that the "Xmas tree" lights were out.
I then checked the console lights and they were out as well, although the power
supply (bottom R side) was still running. So, I powered down the entire
spectrometer, and waited 10-15 minutes, then brought it back up. It came
up OK, with all the lights behaving normally. However, when it came to
locking on a solvent, the lock was behaving erratically. It was tricky
finding the on resonance position on a CDCl3 sample, because the lock signal
would jump from negative to positive. Perhaps this was due to lock phase
being off, especially since when the console would be brought down and then
powered back up, typically the lock phase would change.
I then used a sucrose in h2o/d2o sample, found the initial on-resonance position
at z0=5179, and once locked, it regulated the lock fairly decently. However,
I then proceeded with a 3D gradient autoshimming, which then moved the z0 to
5380-ish, which is quite a bit off from the original value (I often would find
a different z0 after 3D autoshimming, but it would only be 10 or so units off).
And the lock behaved a bit more erratically at this point.
Then I tried the lineshape sample (1%CHCl3 in acetone-d6), the lock behaved
somewhat normal at a higher lock power, but when I brought the lock power
down to 14, the appropriate value, prior to locking the lock level would
bounce from as high as ~83 to as low as ~35. Once locking on the signal,
with z0=1843, the lock bounced a little, and I proceeded to do 1D gradient
autoshimming (having to recable first, since we're detecting 2H in this case).
Unfortunately this autoshimming didn't improve the lock level and resulted
in the lock level bouncing quite a bit when 'regulated'. I decided to
repeat this 1D autoshimming, but after this try, I had to relock, and now z0=2083,
and the lock level is holding in place when 'regulated'. However, when
I went to acquire the spectrum, I found 'lock wiggles' in every peak, and
the peaks were broad.
Finally, I decided to 'reset' the console, in hopes this would improve the
situation, and starting with the sucrose in h2o/d2o sample. Initially
the lock was behaving more or less normally, however when I found on resonance
it was at z0=3903 (wildly different from a few days before with the same
sample). The regulated lock wasn't bad, only fluctuating between 73 and 69.
But then I went and did 3D gradient Autoshimming, but then the x-shim was
taken out of range, AND I had to relock and the new z0=4733. The x1
and y1 shims had been taken way out, so they needed a large manual correction.
However once acquiring the spectrum, the lock was wildly oscillating, between
82% and 69%.
Not sure what else I could try, perhaps bringing the console back down and
then back up, but at some point I'm just wasting my time, so I thought I'd
share this issue with you all to see what you think? Likely I need so
me new boards for the locking and shimming mechanism. And yes, a new
console would be wonderful! ;-)
With gratitude,
Monika
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3785): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3785__;!!PvDODwlR4mBZyAb0!XnwK-16uKozGsrVFtkZpylIf_mu9y4=
ZmcSX3d1gsngi2pHaDLG4Akfpe7ObJK3w5WGgn22tT8jqGfSbaqyWpY8bUNdHE$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121251037=
/7559972__;!!PvDODwlR4mBZyAb0!XnwK-16uKozGsrVFtkZpylIf_mu9y4ZmcSX3d1gsngi2p=
HaDLG4Akfpe7ObJK3w5WGgn22tT8jqGfSbaqyWpY32Cso7P$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Mon Sep 14 2026 - 14:03:43 MST