Re: [AMMRL] Spike Observed in 15N HSQC with Prodigy Probe

From: Spin-Doc <bert.heise_at_spin-doc.net>
Date: Sat, 12 Sep 2026 09:33:05 +0200
Hi,

These may be too obvious but you didn’t write which the “recommended spike-removal
tips and conditions” are…:
- do the spikes go away if you leave all RF pulses but switch off decoupling?
- do they go away if you reduce either 1H or 15N pulse power by some 6 dB?
- did you try a full warmup-cooldown cycle (maybe something got deposited onto the RF coils, freezing in place)?

Not impossible that the probe might need a repair…

Best regards 
Bert

-> Sent via mobile <-

Spin-Doc 
Magnetic resonance equipment, service, training & more
Dr. rer. nat. Bert Heise
Owner, applications scientist
Kiefernweg 13
58239 Schwerte
Germany
Mail: bert.heise_at_spin-doc.net
Phone:+49-171-5135315

Representing MestreLab software, Stelar relaxometers & Barthel HF amplifiers



Am 12.09.2026 um 01:41 schrieb pothula purushotham reddy via groups.io <purush.reddy2008=gmail.com_at_groups.io>:



Hi everyone,

We recently observed spikes in the ^15N HSQC spectrum using our Prodigy probe. I tried the recommended spike-removal tips and conditions, but they did not resolve the issue.

We had observed similar spikes about two years ago. At that time, we subsequently upgraded to TopSpin 4.5 and also upgraded the spectrometer firmware, after which the spikes were no longer observed.

Now, the spikes have appeared again. Does anyone have any suggestions about the possible cause or additional troubleshooting steps we could try?

Any suggestions or advice would be greatly appreciated.


<spike_hsqc.png>


Best regards,

Purush 

_._,_._,_

Groups.io Links:

You receive all messages sent to this group.

View/Reply Onl= ine (#3784) | Reply to Group | Reply to Sender | Mute This Topic | New Topic
Your Subscription | Contact Group Owner |

_._,_._,_
Received on Sat Sep 12 2026 - 14:29:16 MST

This archive was generated by hypermail 2.4.0 : Fri Oct 02 2026 - 02:30:34 MST