Re: [AMMRL] Spike Observed in 15N HSQC with Prodigy Probe

From: Todd Rappe via groups.io <rapp0006=umn.edu_at_groups.io>
Date: Fri, 11 Sep 2026 19:10:10 -0500

We've seen similar on the 13C channel on He-cooled cryoprobes. It seems
like, in our case, contaminants collect on the coil over time. Warming up
the cryoprobe, pumping on it at room temperature for at least 24h and then
cooling it back down generally improves it for us.