Re: [AMMRL] Pulse width variability

From: Atul Srivastava via groups.io <aks275=case.edu_at_groups.io>
Date: Sun, 3 Nov 2024 11:19:38 -0500

Hi Todd,

Adding to Youlin's thought, It could be that the HSQC experiment was being
run with more than the optimal acquisition time. That will make the
decoupler stay on for a long time and can cause the heating of the sample.

Best,
Atul

On Sat, Nov 2, 2024 at 3:58 PM Todd Rappe via groups.io wrote:

> A user in our lab was working optimizing a method and noticed that his 1H
> pulse width was changing after running an experiment. To confirm his
> observation, I ran a series of fast HSQC experiments on the same sample and
> ran pulsecal before each. The results are below:
>
> exp# P1, us
> 2 14.27
> 3 14.85
> 4 14.73
> 5 14.69
> 6 14.87
> 7 15.00
> 8 15.21
> 9 15.57
> 10 15.20
> 11 15.17
> 12 15.06
>
> I checked the tuning between each experiment and the wobble curve did not
> move at all.
>
> I followed up with simply running pulsecal several times in a row to see
> if it is just normal variability in the algorithm, but it gave me a
> consistent value +/- 0.01us each time.
>
> These tests were performed with a 15N-labeled protein sample in a 5-mm
> Shigemi tube on a 600-MHz Bruker Avance NEO w/ TCI cryoprobe.
>
> Does anybody have suggestions what could cause this variability? Thanks
>
>
>


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1822): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1822__;!!PvDODwlR4mBZyAb0!WZffhq6QNGL3126F2Zc5ekXdp27fZO=
IyKeSjVXc6pCiMYnF88Tjnb4ODDvyUegv_XUmOEDBJK0N2tGErsxIT_R6_lSo$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/109357299=
/7559972__;!!PvDODwlR4mBZyAb0!WZffhq6QNGL3126F2Zc5ekXdp27fZOIyKeSjVXc6pCiMY=
nF88Tjnb4ODDvyUegv_XUmOEDBJK0N2tGErsxITFizJfO0$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Sun Nov 03 2024 - 08:20:01 MST

This archive was generated by hypermail 2.4.0 : Sun Dec 01 2024 - 13:25:02 MST