Re: [AMMRL] Pulse width variability

From: clarkson.michael via groups.io <clarkson.michael=gene.com_at_groups.io>
Date: Mon, 04 Nov 2024 09:21:23 -0800

Adding my voice to the sample-heating hypothesis. Salt will increase the
susceptibility to this problem, particularly in a 5 mm tube. Given that you
are running a "fast" HSQC the balance between aq and d1 is something to
consider. If you have an NMR thermometer system you could test this directly;
alternately you could try cutting back 15N decoupling power (the default is
often higher than necessary), decreasing aq, increasing d1, or transferring
the sample to a 3 mm tube (shigemi or otherwise). Any of these should
attenuate or alleviate the problem.


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1828): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1828__;!!PvDODwlR4mBZyAb0!URK69Gd7hHUeMkYNUdBqrfQw3X-Bp-=
UzEK3aJNumLWE_l9AYy6Ww1a5_IoeZZt25EJfCD6pSTggRNLrwF6Qq827_J47T$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/109357299=
/7559972__;!!PvDODwlR4mBZyAb0!URK69Gd7hHUeMkYNUdBqrfQw3X-Bp-UzEK3aJNumLWE_l=
9AYy6Ww1a5_IoeZZt25EJfCD6pSTggRNLrwF6Qq81vSyniA$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Mon Nov 04 2024 - 09:21:30 MST

This archive was generated by hypermail 2.4.0 : Sun Dec 01 2024 - 13:25:02 MST