Re: [AMMRL] Pulse width variability

From: Martha Morton <marthadmorton_at_gmail.com>
Date: Sun, 3 Nov 2024 11:59:47 -0600

Todd,

Did you have temperature and air stability?

Martha

On Sat, Nov 2, 2024 at 2:58 PM Todd Rappe via groups.io wrote:

> A user in our lab was working optimizing a method and noticed that his 1H
> pulse width was changing after running an experiment. To confirm his
> observation, I ran a series of fast HSQC experiments on the same sample and
> ran pulsecal before each. The results are below:
>
> exp# P1, us
> 2 14.27
> 3 14.85
> 4 14.73
> 5 14.69
> 6 14.87
> 7 15.00
> 8 15.21
> 9 15.57
> 10 15.20
> 11 15.17
> 12 15.06
>
> I checked the tuning between each experiment and the wobble curve did not
> move at all.
>
> I followed up with simply running pulsecal several times in a row to see
> if it is just normal variability in the algorithm, but it gave me a
> consistent value +/- 0.01us each time.
>
> These tests were performed with a 15N-labeled protein sample in a 5-mm
> Shigemi tube on a 600-MHz Bruker Avance NEO w/ TCI cryoprobe.
>
> Does anybody have suggestions what could cause this variability? Thanks
>
>
>

--
Martha Morton, PhD
Director of Research Instrumentation, University of Nebraska Lincoln
Safety Co-Chair, Department of Chemistry
Nebraska ACS, Chair 2021
ACS, MWRM Chair 2021


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1823): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1823__;!!PvDODwlR4mBZyAb0!SlLOniE_z3_wpwEyP450Ik-wOdjW7M=
hJxIGZoFOsq0gvXtr4Mk9_4fmT4DbHVPg_Z-z1zMK7EURY7IYp-NiYeQ$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/109357299=
/7559972__;!!PvDODwlR4mBZyAb0!SlLOniE_z3_wpwEyP450Ik-wOdjW7MhJxIGZoFOsq0gvX=
tr4Mk9_4fmT4DbHVPg_Z-z1zMK7EURY7IZZkVuUqQ$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Sun Nov 03 2024 - 09:59:59 MST

This archive was generated by hypermail 2.4.0 : Sun Dec 01 2024 - 13:25:02 MST