Hello Brian,

I think this may have something to do with "GROUP" membership.

What operating system are you using?

My guess is that when you login using a different userid, that the
userid is not a member of the group "nmruser" (or group "nmrsu"). It
doesn't need to be the "Primary Group."  But that new userid must be a
member of nmruser in order to gain full functionality in TopSpin.
Depending on which operating system you are using, there are different
ways to go about adding/changing  "group" memberships.

After you have added the "nmruser" group membership, you probably don't
need to reboot, but I would just to make sure that the new group is
recognized.

Then login and TopShim should work fine for you.

Good luck,
Bob.

On 2025-03-25 2:59 p.m., Brian Antalek via groups.io wrote:
>
>
> I am running TopSpin 3.2 on an AVII 300 MHz system under the nmrsu
> user.  Manual operation and iconnmr automation work fine.
> But, when I switch to a different user, I cannot obtain 1H signal. It
> tunes ok to 1H and locks and shims fine on D2O, but there is no
> acquired signal.
> Before I switched to a different user, I exited TopSpin.  I also
> successfully ran ii when I restarted TopSpin in the other user.
> Has anyone ever seen this?  Does anyone have a remedy?
> thanks
> Brian
>
> Brian Antalek
> Bausch + Lomb, Inc.
> 1400 N. Goodman Street
> Rochester, NY 14609
>
>
> brian.antalek_at_bausch.com <mailto:brian.antalek_at_bausch.com>
>
> (585) 752-8452  mobile
>

--
Dr. Bob Berno
Manager, NMR Facility
McMaster University, ABB-156
1280 Main St. West
Hamilton, ON L8S 4M1
T: 905-525-9140 x24158 or x24686
---------------------------------------------------------------------------
This email contains CONFIDENTIAL information intended for the specific recipient(s).
If you have an accommodation need for a planned meeting, please e-mail me directly and
I will do my best to make appropriate arrangements.
Should you require any materials sent via this e-mail address in an alternate/accessible
format, please let me know.

McMaster University sits on the traditional Territories of the Mississauga and
Haudenosaunee Nations, and within the lands protected by the "Dish With One Spoon"
wampum agreement.


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2169): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2169__;!!PvDODwlR4mBZyAb0!XAIg5BDBcdN9cdLedsqiQ9xyO0fZFn=
93o-Vx1jfMR9x31megf92kkmViqsvJEsCYf8k1LRlsSI8gLpI813w1X6RcIjjr$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/111903971=
/7559972__;!!PvDODwlR4mBZyAb0!XAIg5BDBcdN9cdLedsqiQ9xyO0fZFn93o-Vx1jfMR9x31=
megf92kkmViqsvJEsCYf8k1LRlsSI8gLpI813w1X8EdPtZg$
Group Owner: main+owner_at_ammrl.groups.io
Unsubscribe: https://urldefense.com/v3/__https://ammrl.groups.io/g/main/lea=
ve/12177410/7559972/1125599165/xyzzy__;!!PvDODwlR4mBZyAb0!XAIg5BDBcdN9cdLed=



Received on Wed Mar 26 2025 - 03:06:54 MST

This archive was generated by hypermail 2.4.0 : Tue Apr 01 2025 - 14:21:16 MST