Re: [AMMRL] can get signal under nmrsu but not under a different user

From: Shin, Ronald via groups.io <shinr=uab.edu_at_groups.io>
Date: Wed, 26 Mar 2025 01:42:01 +0000

It could be a permission issue, or could be Topspin not reading in the
parameters properly.

I'd start by creating a new data set, and then executing "getprosol."

After typing in "getprosol," If a simple 1D PROTON experiment shows a 10 us
pulse width and 0 watts of power, then for some reason, Topspin isn't reading
the prosol parameters that were setup.




> From: main_at_ammrl.groups.io On Behalf Of Brian Antalek via groups.io
> Sent: Tuesday, March 25, 2025 2:00 PM
> To: main_at_ammrl.groups.io
> Subject: [AMMRL] can get signal under nmrsu but not under a different user

I am running TopSpin 3.2 on an AVII 300 MHz system under the nmrsu user.
Manual operation and iconnmr automation work fine.
But, when I switch to a different user, I cannot obtain 1H signal. It tunes
ok to 1H and locks and shims fine on D2O, but there is no acquired signal.

Before I switched to a different user, I exited TopSpin. I also successful
ly ran ii when I restarted TopSpin in the other user.

Has anyone ever seen this? Does anyone have a remedy?

thanks
Brian


Brian Antalek
Bausch + Lomb, Inc.
1400 N. Goodman Street
Rochester, NY 14609

http://www.bausch.com
brian.antalek_at_bausch.com<mailto:brian.antalek_at_bausch.com>

(585) 752-8452 mobile



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2170): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2170__;!!PvDODwlR4mBZyAb0!VVUIY6Tl8BQIDoPRmwsIRMO4q-n16w=
uqCSxitD6rRTvP0DPDfzqGnVIKc0ijFbfKYemnj8MeZ0Vs4pja7mtKhawE$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/111903971=
/7559972__;!!PvDODwlR4mBZyAb0!VVUIY6Tl8BQIDoPRmwsIRMO4q-n16wuqCSxitD6rRTvP0=
DPDfzqGnVIKc0ijFbfKYemnj8MeZ0Vs4pja7hyFYSGO$
Group Owner: main+owner_at_ammrl.groups.io
Unsubscribe: https://urldefense.com/v3/__https://ammrl.groups.io/g/main/lea=
ve/12177410/7559972/1125599165/xyzzy__;!!PvDODwlR4mBZyAb0!VVUIY6Tl8BQIDoPRm=



Received on Wed Mar 26 2025 - 03:06:55 MST

This archive was generated by hypermail 2.4.0 : Tue Apr 01 2025 - 14:21:16 MST