Re: [AMMRL] [AMMRL} Topshim freezes issue

From: Spin-Doc <bert.heise_at_spin-doc.net>
Date: Wed, 19 Mar 2025 01:28:00 +0100

Hi,

try to replace the console-internal ethernet switch (if equipped with one)
with another switch - might be that the computer can't find the console.
Depends on which console version you have (older ones don't have a switch).

Best regards
Bert

> Am 18.03.2025 um 19:19 schrieb Brown, Silas via groups.io:
>
> Hello
> We are having a reoccurring issue that pops up occasionally. Topspin wil
> basically freeze, and keep running but nothing will actually run. The
> bottom left of the screen will display getprosol completed successfully,
> and if we try to open the topshim gui and stop the shim it stays frozen. If
> we close topspin and relaunch it works fine for the first run, but the
> second experiment we run will give us the same issue.
>
> The only fix we have come across that resolves the issue is to restart the
> whole console, (PC restart is not enough to resolve issue)
> We are using topspin 3.5.pl7 on a 500 MHz Ulstrasheild.
>
> Any suggestions would be greatly appreciated,
> Silas
> --
> Silas Brown (He/Him), Instrumentation Specialist
> Department of Chemistry
> Williams College
> 150B Morley Science Laboratories
> Williamstown, Ma. 01267
> E: seb14_at_Williams.edu | P: 413.597.3544
>


Magnetic resonance equipment, service, training & more
Selling & buying used NMR equipment (all vendors)
Dr. rer. nat. Bert Heise
Owner, Applications Scientist

Spin-Doc Bert Heise
Kiefernweg 13
58239 Schwerte
Germany
Mobile: +49 171 5135315
Email: bert.heise_at_spin-doc.net
Web: http://www.spin-doc.net

Representing:
Stelar Fast Field Cycling NMR Relaxometers (http://www.stelar.it
MestreLab NMR Software (http://www.mestrelab.com
Barthel HF Amplifiers (http://www.barthel-hf.de


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2155): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2155__;!!PvDODwlR4mBZyAb0!S4DBaugAXMzPZ6r3d0wzYWOYKnSYKm=
HJd0caZkNOYIyTZuMqIOiekW60UfYeVaUm2hLUR1aWm_u8ejKJ1QzeDw$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/111778804=
/7559972__;!!PvDODwlR4mBZyAb0!S4DBaugAXMzPZ6r3d0wzYWOYKnSYKmHJd0caZkNOYIyTZ=
uMqIOiekW60UfYeVaUm2hLUR1aWm_u8ejIS1Ss0TQ$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-



Received on Wed Mar 19 2025 - 04:24:25 MST

This archive was generated by hypermail 2.4.0 : Tue Apr 01 2025 - 14:21:16 MST