Re: [AMMRL] 1H-19F HOESY setup tips?

From: ik54 via groups.io <ik54=cornell.edu_at_groups.io>
Date: Tue, 04 Mar 2025 10:41:29 -0800

Hi Beth,

We use Tim Claridge's semi-selective 1D 19F to 1H method. ( https://doi.org/10.1002/chem.201201577 )
It works well enough to be run under automation in Icon NMR on a 400 MHz AVIII HD with a BBFO probe.
I believe it also requires a QNP switch. It's been a while since I set it up, so I don't recall
the details, but I uploaded our parameter set, pulse program and an recent example to Box:
https://cornell.box.com/s/ui8fksrheuayohvq75pg0fzjtk6ktbpg

Let me know if you have any questions,

Ivan

--
Ivan Keresztes, Ph.D.
Director, NMR Facility
Department of Chemistry and Chemical Biology, Cornell University Baker Laboratory, Ithaca, NY 14853.
phone: (607) 255 0709, fax: (607) 255 4137, email: ik54_at_cornell.edu
(he/him)