Re: [AMMRL] Bruker SampleCase error

From: Michael Groves via groups.io <mgrovesnmr=yahoo.com_at_groups.io>
Date: Thu, 27 Jun 2024 15:08:32 +0000 (UTC)

For what it's worth, I would be surprised if this helped with the samplejet
issue.  The samplecase gets its firmware/operating instructions from the
ELCB while the samplejet is its own thing.  I'm not sure if you've tried
the latest SJ firmware--which admittedly comes with its own gotchas. 
TS3.6 has been pretty stable so that's not an unreasonable workaround.

Cheers,Mike

    On Thursday, June 27, 2024 at 08:23:15 AM MDT, Hansen, Alex wrote:

 Thanks for that! I would be interested in the file as this has been plaguing
 one of our 600s.

Do you happen to know if this also fixes issues with SampleJets using IconNMR?
We have a problem switching from Icon and then running samples normally
(with 'sx #') on the new TopSpin 3.7. Proposed workarounds have not been
consistent, so we've downgraded to a 3.6 version that doesn't experience the
error.

Alex

Dr Alexandar L Hansen
Lead Research Scientist
CCIC & Gateway NMR Facility
R224 CBEC, 151 W Woodruff Ave, Columbus, OH 43210
614.292.7133 Office / 716.341.8476 Mobile
hansen.434_at_osu.edu ccic.ohio-state.edu

Pronouns: he/him/his

“Who's ready to make some science?â€=
� – Cave Johnson

 
> From: main_at_ammrl.groups.io on behalf of Perry Pellechia
> Sent: Thursday, June 27, 2024 8:54 AM
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: Re: [AMMRL] Bruker SampleCase error  

I am reviving this tread so the archives are complete.  Bruker has release
new ELCB firmware (elcb_firmware_240617) that addresses the issue with
slight contact of the sample height (red) bar on the SampleCase causing it=
to go into fault mode.  The new firmware will immediately clear the
error and not interrupt the system.

I have been asked not to post the file but I can send it to anyone that asks
(who are not in sanctioned countries).  You could also request the file
from your Bruker service center.  The firmware has been installed and
running on my AVIII-HD, running Topspin 3.7.0 on Win 11.  Check with
Bruker to see if other consoles/TS versions are supported.

A special thanks to Dimitris Raptis who got the R&D team to make this change.  
Thanks everyone who might have made the same request to Bruker to make this change.

--
Perry.
--

Perry J. Pellechia
Director of NMR ServicesResearch Professor
University of South Carolina
Department of Chemistry and Biochemistry


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1375): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1375__;!!PvDODwlR4mBZyAb0!RA88EkF4-KMUBhQOxEkoq6phDGveCw=
_UJzk29HrlFn8PXdXkd7zDrOrYOI6euQ3r5uYA8CG-JBeT77L_hJ2IU9jqdFUU$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/105565488=
/7559972__;!!PvDODwlR4mBZyAb0!RA88EkF4-KMUBhQOxEkoq6phDGveCw_UJzk29HrlFn8PX=
dXkd7zDrOrYOI6euQ3r5uYA8CG-JBeT77L_hJ2IU3Z_1Tvj$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Thu Jun 27 2024 - 08:31:53 MST

This archive was generated by hypermail 2.4.0 : Mon Jul 01 2024 - 12:45:11 MST