Dear Julian,
We experienced this error several times on our 500 Av III where a lot of
pulse programs a created. Bruker gave us this solution : delete the file
(or put it somewhere else) called "pulsprogs-1.xml" which is in
.../topspinxxx/prog/curdir/nmr/db/Local/pp This should solve the problem.
Then you could think about doing some cleaning in you own pulse programs :)
Hope this may help !
Cheers
*Aurélie BERNARD*
Ingénieure RMN
Laboratoire CEISAM
UMR CNRS 6230
Tel : 02 51 12 54 11 (interne 45 54 11)
Mail : aurelie.bernard_at_univ-nantes.fr
aurelie.bernard_at_cnrs.fr
Université de Nantes
2 chemin de la Houssinière
bâtiment 22, BP 92 208
44322 Nantes Cedex 3
Bureau : 014
https://ceisam.univ-nantes.fr
*Adoptez l'éco-attitude : n'imprimez ce mail que si cela est vraimentnécessaire*
Le 19/06/2024 à 16:44, Julian Holzinger a écrit :
>
> Dear all,
>
> We recently ran into a java heap space problem (see attachment).
> The error pops up when we try to open our pulse sequence library in
> TopSpin.
> Apparently (at least according to a quick Google search) one solution
> might be to increase the Java heap space. However, I have no idea how
> to do this on our system.
> Have any of you much more experienced ladies and gents encountered the
> same problem?
>
> Many thanks and all the best!
> Julian
>
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1374): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1374__;!!PvDODwlR4mBZyAb0!QHI1G-rr4f-cIiD6Jzr0myyyclhNR0=
dI6z4Iy9-dnOgNimToRTc1_3itYq5YUb5C6XiV-mHF6qg8QlATfA0XsSa66qWMAwvv$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/106785294=
/7559972__;!!PvDODwlR4mBZyAb0!QHI1G-rr4f-cIiD6Jzr0myyyclhNR0dI6z4Iy9-dnOgNi=
mToRTc1_3itYq5YUb5C6XiV-mHF6qg8QlATfA0XsSa66u32MZLU$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Thu Jun 27 2024 - 07:23:23 MST