Hello!

Von meinem iPad gesendet

Am 21.06.2024 um 21:25 schrieb Spin-Doc <bert.heise_at_spin-doc.net>:

Hi Luke,

there may be another possible solution (bit more drastic, but here you go): In case the installation of the new NMR system isn't too far away anymore you could also just quench the old magnet yourself - being unshielded it's rather unlikely that it will find a new home (unless you do have such plans for it of course). Quenching a magnet - specifically an older one which has a readily accessible vacuum valve - is shockingly easy…

Yes - this is working fine. There is a closed cap on before the valve to the vacuum chamber. We opened the cap - closed it and opened the valve.
Closed the valve - opened the cap vented the small volume there and then closed this part with a cap and opened the valve to the vacuum once more again. You have to do this 10 or more times.

This takes time - but the vacuum is with this method slowly broken down. after a while the quenching vents are opened by the magnet itself - the quench is happening now. 

In our case - we tried to recover as much as possible of the liquid helium. So we tried to vent the main stuff in to our recovery system. Before all this - the main bladder was completely emptied and during the quench the recovery system went on maximum. Even we tried to put more into our recovery system and put a screw driver over the quenching vents and press this one a bit down - venting into the recovery hose. This was the final end of my 9.4 Tesla Oxford magnet. 

Bert is right - its working.

Yours sincerecly,

Ulrich Haunz


_._,_._,_

Groups.io Links:

You receive all messages sent to this group.

View/Reply Online (#1370) | Reply To Group | Reply To Sender | Mute This Topic | New Topic
Mute #heliumrecovery
Your Subscription= | Contact Group Owner |

_._,_._,_
Received on Wed Jun 26 2024 - 05:58:07 MST

This archive was generated by hypermail 2.4.0 : Mon Jul 01 2024 - 12:45:11 MST