Re: [AMMRL] Negative 13C signal from zgpg30 experiment.

From: Anklin, Clemens via groups.io <clemens.anklin=bruker.com_at_groups.io>
Date: Tue, 28 Jul 2026 18:23:32 +0000

Hi Beth

if you zero fill the data to get better resolution you can see that this peak
is out of phase and not fully negative and very narrow. This looks like a
spurious signal to me. That far in the spectrum it can not be a folded or
aliased peak. it is also curious that it appears at exactly 125.78000000 MHz.
This is a frequency used by aviation communications.

Make sure all BNC, N-type and SMA connectors are tight. Then check if the
signal is still there.

With best regards

Clemens


On 7/28/2026 12:18 PM, Beth Moscato via groups.io wrote:

Hi everyone –

A student just acquired a 13C spectrum that has me stumped. Using standard
13C parameters (zgpg30 pulse sequence, d1 + aq = 5 sec, ns = 4k), he has
a spectrum where all signals are phased positively except one at 176ppm which
is negative. An adjacent carbonyl has a positive phase, as do peaks around 0 ppm.

The student insists that this is a signal from the target molecule and not
an artifact.

This is acquired on a largely well behaved 500MHz AVANCE III system with a
BBO probe. I have seen spurious tiny signals in the past from 31P and molybdenum,
both of which I’ve attributed to oil mist deposits in the system from
previous compressors ( https://www.ammrl.org/archives/May-99/0047.html ), but
I’ve not encountered something with an intensity similar to other signals.
There’s another 500MHz in the room (both unshielded), but their frequencies
are offset enough that I’ve never observed crosstalk.

Has anyone observed anything similar? I’ve attached the jdx file for
reference.

~~ Beth Moscato, Temple University



--
==========================
==========================
====================
Clemens Anklin Ph.D. Vice President
Bruker BioSpin NMR Applications & Training
15 Fortune Drive NEW Phone: 978-313-5NMR(5667)
Billerica MA 01821 bluesky: _at_canklin.bsky.social
web: https://urldefense.com/v3/__http://www.bruker.com__;!!PvDODwlR4mBZyAb0=
!Sdc0x8zvjvTTspZ_rocqe2GzRfadgvSL5OmtijiS0KSQf7dtuRJlcrtBMJK1bQrge9GXrjD5MZ=
hQef3zDBBQmQJuQ7af$ <https://urldefense.com/v3/__http://www.bruker.com__;!!=
PvDODwlR4mBZyAb0!Sdc0x8zvjvTTspZ_rocqe2GzRfadgvSL5OmtijiS0KSQf7dtuRJlcrtBMJ=
K1bQrge9GXrjD5MZhQef3zDBBQmQJuQ7af$ > e-mail: clemens.anklin_at_bruker.com<ma=
ilto:clemens.anklin_at_bruker.com>
==========================
==========================
====================



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3618): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3618__;!!PvDODwlR4mBZyAb0!Sdc0x8zvjvTTspZ_rocqe2GzRfadgv=
SL5OmtijiS0KSQf7dtuRJlcrtBMJK1bQrge9GXrjD5MZhQef3zDBBQmQVBRpv0$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120486664=
/7559972__;!!PvDODwlR4mBZyAb0!Sdc0x8zvjvTTspZ_rocqe2GzRfadgvSL5OmtijiS0KSQf=
7dtuRJlcrtBMJK1bQrge9GXrjD5MZhQef3zDBBQmTGGXaxc$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue Jul 28 2026 - 11:26:46 MST

This archive was generated by hypermail 2.4.0 : Sat Aug 01 2026 - 17:50:19 MST