Hi,
if it's an external signal coming from some kind of radio source it should
be completely independent of your spectrometer's RF pulses - so try to set
the 13C pulse power to zero/tiny power level and acquire the same spectrum
like this - if the spike is still present then it's confirmed as external.
If that turns out to be the case then there are - besides Clemens' good
suggestions of course - only rather drastic measures that can help like RF
shielding the NMR room or moving the magnet field frequency...
Cheers
Bert
> Am 28.07.2026 um 20:23 schrieb Anklin, Clemens via groups.io:
>
> Hi Beth
>
> if you zero fill the data to get better resolution you can see that this peak
> is out of phase and not fully negative and very narrow. This looks like a
> spurious signal to me. That far in the spectrum it can not be a folded or
> aliased peak. it is also curious that it appears at exactly 125.78000000 MHz.
> This is a frequency used by aviation communications.
>
> Make sure all BNC, N-type and SMA connectors are tight. Then check if the
> signal is still there.
>
> With best regards
>
>
> Clemens
>
>
> On 7/28/2026 12:18 PM, Beth Moscato via groups.io wrote:
>>
>> Hi everyone
>>
>> A student just acquired a 13C spectrum that has me stumped. Using standard
>> 13C parameters (zgpg30 pulse sequence, d1 + aq = 5 sec, ns = 4k), he
>> has a spectrum where all signals are phased positively except one at 176ppm
>> which is negative. An adjacent carbonyl has a positive phase, as do peaks
>> around 0 ppm.
>>
>> The student insists that this is a signal from the target molecule and not an artifact.
>>
>> This is acquired on a largely well behaved 500MHz AVANCE III system with a BBO
>> probe. I have seen spurious tiny signals in the past from 31P and molybdenum,
>> both of which I’ve attributed to oil mist deposits in the system from
>> previous compressors ( https://www.ammrl.org/archives/May-99/0047.html ),
>> but I’ve not encountered something with an intensity similar to other
>> signals. There’s another 500MHz in the room (both unshielded), but their
>> frequencies are offset enough that I’ve never observed crosstalk.
>>
>> Has anyone observed anything similar? I’ve attached the jdx file for reference.
>>
>> ~~ Beth Moscato, Temple University
>
Magnetic resonance equipment, service, training & more
Selling & buying used NMR equipment (all vendors)
Dr. rer. nat. Bert Heise
Owner, Applications Scientist
Spin-Doc Bert Heise
Kiefernweg 13
58239 Schwerte
Germany
Mobile: +49 171 5135315
Email: bert.heise_at_spin-doc.net Web: http://www.spin-doc.net
Representing:
Stelar Fast Field Cycling NMR Relaxometers (
http://www.stelar.it )
MestreLab NMR Software (
http://www.mestrelab.com )
Barthel HF Amplifiers (
http://www.barthel-hf.de )
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3622): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3622__;!!PvDODwlR4mBZyAb0!STQEYxQntM0q59M_WAQNG5h7ajmxvN=
HDn8nSc7BzkWsausMUCdwztRygyg3IYwP3jA5nOOfB7am-RmqHgcyURw$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120486664=
/7559972__;!!PvDODwlR4mBZyAb0!STQEYxQntM0q59M_WAQNG5h7ajmxvNHDn8nSc7BzkWsau=
sMUCdwztRygyg3IYwP3jA5nOOfB7am-Rmp3ztu4nQ$
Group Owner: main+owner_at_ammrl.groups.io
Unsubscribe: https://urldefense.com/v3/__https://ammrl.groups.io/g/main/lea=
ve/12177410/7559972/1125599165/xyzzy__;!!PvDODwlR4mBZyAb0!STQEYxQntM0q59M_W=
AQNG5h7ajmxvNHDn8nSc7BzkWsausMUCdwztRygyg3IYwP3jA5nOOfB7am-Rmpm2ICVTg$ [wa=
lt_at_hawaii.edu]
-=-=-=-=-=-=-=-=-=-=-=-
Received on Wed Jul 29 2026 - 04:17:56 MST