Re: [AMMRL] Guidance on Reducing LN₂ Consumption for the Prodigy Probe During Weekends

From: Spin-Doc <bert.heise_at_spin-doc.net>
Date: Wed, 29 Jul 2026 00:52:33 +0200

Hi Purush,

have you attempted to ask around chemical etc. companies in your vicinity
whether they might have some samples to fill the weekend (hoping you're
having some type of autosampler allowing to run some stuff overnight/over
the weekend)…? You might attract some cherished third-party money
in the process while keeping the NMR busy ;-)

Cheers
Bert

> Am 28.07.2026 um 08:14 schrieb Rentsch, Daniel via groups.io:
>
> Hi Purush
>
> It is not recommendable to warmup/cooldown the Prodigy probe too often since
> every warming/cooling cycle impacts the vacuum and electronics part of the probe.
> So I would NOT warm the system up for 1 weekend only…..
>
> If you have a gap of 2 or more week in between use, maybe warming up makes sense.
> To restart the system, you should first turn on the "pump CPR vacuum in warm state"
> function for at least 2 days and then you can make use of the "timer" function in
> prodigy menu, where you can define the start of the next cooldown (of course the
> dewar must be filled with liquid N2).
>
> Best
> Dani
>
> Von: main_at_ammrl.groups.io Im Auftrag von pothula purushotham reddy via groups.io
> Gesendet: Donnerstag, 23. Juli 2026 16:30
> An: main_at_ammrl.groups.io
> Betreff: [AMMRL] Guidance on Reducing LN₂ Consumption for the Prodigy Probe During Weekends
>
> Hi everyone,
>
> We recently installed and upgraded the Avance NEO 400 MHz spectrometer with a Prodigy probe.
>
> Since the spectrometer is generally not used during weekends and university holidays,
> we would like to minimize liquid nitrogen consumption during these periods. Could you
> please advise on the recommended procedure or best practices for safely operating the
> Prodigy platform in a standby or reduced-consumption mode when the instrument will be
> idle for 48–72 hours?
>
> We would also appreciate any guidelines for handling the Prodigy platform during extended
> periods of non-use, while ensuring that the probe can be cooled down and returned to
> operational status quickly on Monday morning.
>
> Thank you for your assistance. We look forward to your recommendations.
>
> Best regards,
> Purush
>


Magnetic resonance equipment, service, training & more
Selling & buying used NMR equipment (all vendors)
Dr. rer. nat. Bert Heise
Owner, Applications Scientist

Spin-Doc Bert Heise
Kiefernweg 13
58239 Schwerte
Germany
Mobile: +49 171 5135315
Email: bert.heise_at_spin-doc.net Web: http://www.spin-doc.net

Representing:
Stelar Fast Field Cycling NMR Relaxometers (http://www.stelar.it )
MestreLab NMR Software (http://www.mestrelab.com )
Barthel HF Amplifiers (http://www.barthel-hf.de )



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3621): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3621__;!!PvDODwlR4mBZyAb0!QJC73bXjFFutntuKZoh3XhHRPxcZHC=
QV7XmmO1nJ62jELF8AevWvn8OM1f9ehyGSDfkSmZwPzVKn-3aoHtrGGg$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120474322=
/7559972__;!!PvDODwlR4mBZyAb0!QJC73bXjFFutntuKZoh3XhHRPxcZHCQV7XmmO1nJ62jEL=
F8AevWvn8OM1f9ehyGSDfkSmZwPzVKn-3aCBblXLQ$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-



Received on Tue Jul 28 2026 - 19:30:48 MST

This archive was generated by hypermail 2.4.0 : Sat Aug 01 2026 - 17:50:19 MST