Re: [AMMRL] Guidance on Reducing LN₂ Consumption for the Prodigy Probe During Weekends

From: Rentsch, Daniel via groups.io <Daniel.Rentsch=empa.ch_at_groups.io>
Date: Tue, 28 Jul 2026 06:14:33 +0000

Hi Purush

It is not recommendable to warmup/cooldown the Prodigy probe too often since
every warming/cooling cycle impacts the vacuum and electronics part of the probe.

So I would NOT warm the system up for 1 weekend only…..

If you have a gap of 2 or more week in between use, maybe warming up makes
sense.

To restart the system, you should first turn on the "pump CPR vacuum in warm
state" function for at least 2 days and then you can make use of the "timer"
function in prodigy menu, where you can define the start of the next cooldown
(of course the dewar must be filled with liquid N2).

Best

Dani

> Von: main_at_ammrl.groups.io Im Auftrag von pothula purushotham reddy via groups.io
> Gesendet: Donnerstag, 23. Juli 2026 16:30
> An: main_at_ammrl.groups.io
> Betreff: [AMMRL] Guidance on Reducing LN₂ Consumption for the Prodigy Probe During Weekends

Hi everyone,

We recently installed and upgraded the Avance NEO 400 MHz spectrometer with
a Prodigy probe.

Since the spectrometer is generally not used during weekends and university
holidays, we would like to minimize liquid nitrogen consumption during these
periods. Could you please advise on the recommended procedure or best practices
for safely operating the Prodigy platform in a standby or reduced-consumption
mode when the instrument will be idle for 48–72 hours?

We would also appreciate any guidelines for handling the Prodigy platform
during extended periods of non-use, while ensuring that the probe can be
cooled down and returned to operational status quickly on Monday morning.

Thank you for your assistance. We look forward to your recommendations.

Best regards,
Purush



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3615): https://ammrl.groups.io/g/main/message/3615
Mute This Topic: https://groups.io/mt/120474322/7559972
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue Jul 28 2026 - 06:44:10 MST

This archive was generated by hypermail 2.4.0 : Sat Aug 01 2026 - 17:50:19 MST