Hi everyone –
A student just acquired a 13C spectrum that has me stumped. Using standard
13C parameters (zgpg30 pulse sequence, d1 + aq = 5 sec, ns = 4k), he has
a spectrum where all signals are phased positively except one at 176ppm which
is negative. An adjacent carbonyl has a positive phase, as do peaks around 0 ppm.
The student insists that this is a signal from the target molecule and not
an artifact.
This is acquired on a largely well behaved 500MHz AVANCE III system with a
BBO probe. I have seen spurious tiny signals in the past from 31P and molybdenum,
both of which I’ve attributed to oil mist deposits in the system from
previous compressors (
https://www.ammrl.org/archives/May-99/0047.html), but
I’ve not encountered something with an intensity similar to other signals.
There’s another 500MHz in the room (both unshielded), but their frequencies
are offset enough that I’ve never observed crosstalk.
Has anyone observed anything similar? I’ve attached the jdx file for
reference.
~~ Beth Moscato, Temple University
-=-=-=-=-=-=-=-=-=-=-=-