Re: [AMMRL] J Young Tubes getting stuck in Bruker 400 with CASE autosampler

From: Ezell, Ed via groups.io <eezell=unmc.edu_at_groups.io>
Date: Mon, 6 Jul 2026 21:10:02 +0000

Hey Guys, I generally don't have anything to add to the excellent suggested
solutions from the community, but just this morning I was running some VT
acquisitions and used the ceramic spinner for the first time after the SampleCase
installation, and it was too heavy! If you are quick about it and get to the
top while eject air is still on, a light bit of finger pressure on the sample
carriage to make the mating with the upper bore a bit tighter temporarily and
the sample pops right up. My bit of " AI" (animal intelligence).

EdEz


Ed Ezell

NMR Research Manager

Eppley Institute
University of Nebraska Medical Center
Eppley Science Hall (ESH) 1015 | Omaha, NE 68198-6805
402.559.4174

eezell_at_unmc.edu

 UNMC<http://www.unmc.edu/> | Facebook<https://www.facebook.com/unmcedu | X<https://x.com/unmc>
Instagram<https://instagram.com/iamunmc> | YouTube< https://www.youtube.com/user/UNMCEDU >
Bluesky<https://bsky.app/profile/unmc.bsky.social>

________________________________
I> From: main_at_ammrl.groups.io on behalf of Michael Groves via groups.io
> Sent: Monday, July 6, 2026 10:25 AM
> To: main_at_ammrl.groups.io
> Subject: Re: [AMMRL] J Young Tubes getting stuck in Bruker 400 with CASE autosampler


We're 100% fine with sharing the BSMS service login. The username must be
at least 3 characters but it doesn't matter what you use. I've started telling
people to just use nmrsu because it saves me having to type the above.
.. password is 1964

And with the samplecase, I believe if you change the lift parameters you have
to reset the sample changer to apply them. If it actually does anything.

Cheers,
Mike


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3546): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3546__;!!PvDODwlR4mBZyAb0!SUdMNtDb7RFO4JnjNC6K_n5JN3VHZH=
vbakiP2QxPJUzb05bfS5m-AMKzUfbwIikgwzMITyKKbemr3-BmJ91nt2b968Y$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120139056=
/7559972__;!!PvDODwlR4mBZyAb0!SUdMNtDb7RFO4JnjNC6K_n5JN3VHZHvbakiP2QxPJUzb0=
5bfS5m-AMKzUfbwIikgwzMITyKKbemr3-BmJ91ntKVHwS0$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Mon Jul 06 2026 - 15:00:56 MST

This archive was generated by hypermail 2.4.0 : Sat Aug 01 2026 - 17:50:19 MST