Hi Sebastian,
We capture all of our helium and run the compressors during transfers, so what
you are suggesting is certainly possible. We have two bags (approximately
17m^3 each) and two CryoMech HRCP compressors attached (one to each bag).
During large transfers on our big magnets where we transfer ~220L (~180L of liquid
transferred), we utilize the full volume of the bags while running two compressors.
I like to assume around 20% flashed gas during large transfers, and about 15% on
small transfers. On small transfers around ~50L for small magnets, we
typically don't run a compressor during the fill, and it raises each bag by only
about 25% or less. I'd have to crunch some numbers to verify, but my
intuition tells me that a 12m^3 bag and a running compressor should be adequate
for capturing all the gas from a 80-90L fill at typical transfer rates.
When we know the full bag volume will be necessary during a fill, our HRCP
compressor allows us to bypass the sensors and run it manually.
We empty the bags before starting a large transfer. During normal operation,
our laser typically triggers the compressors to turn on around ~70% of their
volume and stop around ~10% or so (to avoid vacuuming our line). During
large fills, we manually start the compressors as soon as liquid is being
transferred.
I realize this doesn't answer all your questions, but hopefully it gives you
some useful information! Feel free to reach out if I can help with
anything.
Best,
Samrat Amin
Magnetic Resonance Research Center
Arizona State University
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2603): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2603__;!!PvDODwlR4mBZyAb0!XNLJ8ZptFAq72z3EJoSXtHbx-icuIt=
zh1Sbtzpb4wHypbmMDZXZ3VKbvOYfl0NghHhaATbkokYVH1gxqMmw-6vm8JJOg$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/114436217=
/7559972__;!!PvDODwlR4mBZyAb0!XNLJ8ZptFAq72z3EJoSXtHbx-icuItzh1Sbtzpb4wHypb=
mMDZXZ3VKbvOYfl0NghHhaATbkokYVH1gxqMmw-6hx7-lDs$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Tue Jul 29 2025 - 12:33:47 MST