Re: [AMMRL] Coolant for He Compressors

From: Todd Rappe via groups.io <rapp0006=umn.edu_at_groups.io>
Date: Tue, 29 Jul 2025 14:29:20 -0500

Our chilled water loop is similar to the one Alex describes at NMRFAM. It
had been problem-free for the past 14 years until this past winter we had a
catastrophic failure. Here's my cautionary tale. Over the Christmas
holiday, one of our helium compressors failed in a manner that leaked oil
into our chilled water lines. The contaminated chilled water ultimately
caused the failure of three other compressors on the loop. I will note that
all the compressors that failed all had over 100,000 hours so we knew that
replacement was somewhat imminent but we weren't expecting to have to
replace them all at the same time. Four "younger" compressors survived. We
contracted with a local company to service our chilled water system and
clean out the oil. Everything is back up and running now. The moral of this
story: think about your replacement schedule for helium compressors and
there may be some value in having each cryocooler on its own isolated
cooling system (we still don't). Also, I believe the initial compressor
failure was caused by a corroded/clogged quick-connect fitting on the
chilled water input to the compressor. I've replaced all of the
Bruker-provided quick-connect fittings on our compressors with wider
diameter hose barbs to hopefully prevent such clogging in the future.

_____________________________________
Todd Rappe
Minnesota NMR Center <nmr.umn.edu>
612-625-9147 (office)
651-321-TODD (rings office and mobile)
612-624-8892 (magnet lab)


On Tue, Jul 29, 2025 at 1:52 PM Alex Paterson via groups.io wrote:

> Hi Luke,
>
> Re 2, is it possible for you to split the difference and have a heat
> exchanger between your building chilled and your process loop? That’s how
> we do it at NMRFAM.
>
> Our process chilled water loop is closed and exclusively serves our
> equipment (approx. 14 helium compressors). It has dedicated redundant
> lead-lag circulation pumps. Physical Plant does periodic maintenance on the
> pumps and the loop itself.
>
> The process chilled water loop rejects heat to the campus chilled water
> system via a pair of heat exchangers.
>
> It’s a non-trivial amount of space, but we’ve found this arrangement to be
> highly effective. Keeping our process loop closed allows us to keep the
> working fluid within helium compressor specifications (no algae!) while
> reducing the amount and complexity of equipment we need to maintain.
>
> Best,
>
> Alex
>
> Alex Paterson
> The National Magnetic Resonance Facility at Madison (NMRFAM)
> University of Wisconsin-Madison
>
> alpaterson_at_wisc.edu
>
>
>
> *From:* main_at_ammrl.groups.io *On Behalf Of *Fulton, Luke via groups.io
> *Sent:* Tuesday, July 29, 2025 11:27 AM
> *To:* main_at_ammrl.groups.io
> *Subject:* [AMMRL] Coolant for He Compressors
>
> Hi all,
>
> I'm planning end-of-life transition for our chilled water recirculator
> which services two helium compressor cold heads. The main unit sits in a
> hallway just outside the lab, and the building's design doesn't allow for
> putting it anywhere else.
>
> 1) Any recommendations for vendor or chiller type? I'm also interested in
> general quotes (feel free to message me privately with any insights on what
> we might expect).
>
> 2) Some in my department are pushing for integrating into the chemistry
> building's existing chilled water network. I've not heard of labs choosing
> this approach, and my impression is endless complications from algae and
> other sundry building issues. Anyone seen success with this? Or is it one
> of those 39-and-a-half-foot-pole topics?
>
> Kind regards,
> Luke
>
>
> Luke Fulton, PhD CHEM BLDG R003
> NMR Core Facility Director Unit 3060
> COR2E & Department of Chemistry 55 N Eagleville Road
> University of Connecticut Storrs, CT 06279
>
> email: fko24003_at_uconn.edu
> (alias): luke.fulton_at_uconn.edu
>
>
>
>


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2602): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2602__;!!PvDODwlR4mBZyAb0!SOIAHLYPn7K5GpTvxb1CU5-dpCeXib=
gM0VdG6cr29d-ItKEYF_-PXA5-d0u3qdGp64tufrP8OxsmnS4IyGG7acnr1zQX$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/114434070=
/7559972__;!!PvDODwlR4mBZyAb0!SOIAHLYPn7K5GpTvxb1CU5-dpCeXibgM0VdG6cr29d-It=
KEYF_-PXA5-d0u3qdGp64tufrP8OxsmnS4IyGG7aYBwXsgI$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue Jul 29 2025 - 12:33:40 MST

This archive was generated by hypermail 2.4.0 : Sat Aug 02 2025 - 14:13:11 MST