Hello everyone,

I would like to modify my helium recovery system so that I can collect helium
during the filling procedure. Due to the relatively limited space, I have a
few questions about balloon size and compressor performance and I would be
very grateful for any experience from colleagues so that I can purchase the
right equipment.

- How much helium will be produced with a fill of 80-90 liters? I calculated
for 80 liters with a fill loss of 15% that max. 16 m³ will be released during
the fill (assuming the cold replaced gas will warm up from 4 K to 298 K). What
are the real numbers? Are there any empirical values?

- I probably only have space for a balloon of 10-12 m³. Do you have a setting
where the compressor starts during filling and compensates for a balloon that is
'too small'? A colleague told me, that it is possible, but I don't want to
miscalculate and it would be very helpful to hear about other cases. If you have
such a setting, what size balloon do you have? What is the compressor's flow rate?
At what percentage of the balloon volume does the compressor activate?

Best regards,
Sebastian


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2598): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2598__;!!PvDODwlR4mBZyAb0!TIBykXo_xAu_ak3SHeyFHr-REP5I2L=
A_IPblUQv_xLRR2BMNUp2l2WPv8B5_2d_lF992SUL3F22ZhID_w_BJ9-4I6igr$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/114436217=
/7559972__;!!PvDODwlR4mBZyAb0!TIBykXo_xAu_ak3SHeyFHr-REP5I2LA_IPblUQv_xLRR2=
BMNUp2l2WPv8B5_2d_lF992SUL3F22ZhID_w_BJ923lwT6s$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue Jul 29 2025 - 11:52:43 MST

This archive was generated by hypermail 2.4.0 : Sat Aug 02 2025 - 14:13:11 MST