Re: [AMMRL] 19F and 1H single channel HOESY (Varian)

From: Spin-Doc <bert.heise_at_spin-doc.net>
Date: Tue, 29 Jul 2025 00:10:13 +0200

Hi Shawn,

you should normally have a "19F decouple" autocalibration under "F19 obs/dec"
which should give you pwx/pwxlvl/pwxlvl_cf for 19F. The WURST40 sequence is
then generated on-the-fly based on this calibration.

For the 1H decoupling shape (should also be a WURST40, not WALTZ) a correct
calibration of pp/pplvl and compression factor is required.

Cheers
Bert

> Am 28.07.2025 um 23:27 schrieb Shawn Wagner via groups.io:
>
> Thanks Bert,
>
> You are correct! I can use the 1H-19F sequences and disregard the chan2 label.
> I was able to acquire 19F with 1H decoupling after adjusting the 1H decoupling
> power slightly. I could also get a 1H spectrum with 19F decoupling if I used a
> WALTZ decoupling centered on the 19F nuclei frequency, but the WURST default
> sequence was ineffective.
>
> Any insights as to anything special that needs to be done to make the WURST
> decoupling work would be helpful. Is that something that has to be added to
> the probe file for the sequence powers to calculate correctly?
>
> Shawn
>
>
> From: main_at_ammrl.groups.io on behalf of Baas, Sander via groups.io
> Sent: Thursday, July 24, 2025 2:16 PM
> To: main_at_ammrl.groups.io
> Subject: Re: [AMMRL] 19F and 1H single channel HOESY (Varian)
>
> Hi Shawn,
>
> Thanks for clarifying.
>
> I think Bert and I both mean you will get both the 1H and 19F pulses out
> of the amplifier high band channel. The VNMRS console has an internal
> splitter/combiner which can route PTS channels 1 and 2 towards a single
> amplifier high band channel. If the probe you use has a single coil/RF
> connector for 1H and 19F, which is tuned for both frequencies, your
> experiment should run fine. The VNMRS knows it doesn't have a second high
> band channel and will route all 1H and 19F pulses through Amp 1 Channel 1
> (high band), even when running 2D 1H/19F experiments.
>
> In this case, there is a difference in PTS channels and amplifier channels.
> PTS will use channel 1 and 2, the amplifier will only use channel 1 (high band).
>


Magnetic resonance equipment, service, training & more
Selling & buying used NMR equipment (all vendors)
Dr. rer. nat. Bert Heise
Owner, Applications Scientist

Spin-Doc Bert Heise
Kiefernweg 13
58239 Schwerte
Germany
Mobile: +49 171 5135315
Email: bert.heise_at_spin-doc.net
Web: http://www.spin-doc.net

Representing:
Stelar Fast Field Cycling NMR Relaxometers (http://www.stelar.it)
MestreLab NMR Software (http://www.mestrelab.com)
Barthel HF Amplifiers (http://www.barthel-hf.de)



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2587): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2587__;!!PvDODwlR4mBZyAb0!Q2rW5Bm8C7Uh6HIUKQf8BIMvD-HFQ3=
JO-R7705bH8qvtQHQN9cFAOzUMS5TYsQT6mCQTsz0-ldwFr28ED7anvQ$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/114278283=
/7559972__;!!PvDODwlR4mBZyAb0!Q2rW5Bm8C7Uh6HIUKQf8BIMvD-HFQ3JO-R7705bH8qvtQ=
HQN9cFAOzUMS5TYsQT6mCQTsz0-ldwFr29q3ezH8A$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-



Received on Mon Jul 28 2025 - 17:31:27 MST

This archive was generated by hypermail 2.4.0 : Sat Aug 02 2025 - 14:13:11 MST