Re: [AMMRL] 19F and 1H single channel HOESY (Varian)

From: Shawn Wagner via groups.io <shawn.wagner=usc.edu_at_groups.io>
Date: Mon, 28 Jul 2025 21:27:37 +0000

Thanks Bert,

You are correct! I can use the 1H-19F sequences and disregard the chan2 lab=
el. I was able to acquire 19F with 1H decoupling after adjusting the 1H dec=
oupling power slightly. I could also get a 1H spectrum with 19F decoupling =
if I used a WALTZ decoupling centered on the 19F nuclei frequency, but the =
WURST default sequence was ineffective.

Any insights as to anything special that needs to be done to make the WURST=
 decoupling work would be helpful. Is that something that has to be added t=
o the probe file for the sequence powers to calculate correctly?

Shawn


________________________________
From: main_at_ammrl.groups.io <main_at_ammrl.groups.io> on behalf of Baas, Sander=
 via groups.io <sander.baas=wur.nl_at_groups.io>
Sent: Thursday, July 24, 2025 2:16 PM
To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
Subject: Re: [AMMRL] 19F and 1H single channel HOESY (Varian)

Hi Shawn, Thanks for clarifying. I think Bert and I both mean you will get =
both the 1H and 19F pulses out of the amplifier high band channel. The VNMR=
S console has an internal splitter/combiner which can route PTS channels 1 =
and 2 towards a

Hi Shawn,

Thanks for clarifying.

I think Bert and I both mean you will get both the 1H and 19F pulses out of
the amplifier high band channel. The VNMRS console has an internal splitter/combiner
which can route PTS channels 1 and 2 towards a single amplifier high band channel.
If the probe you use has a single coil/RF connector for 1H and 19F, which is tuned
for both frequencies, your experiment should run fine. The VNMRS knows it doesn't
have a second high band channel and will route all 1H and 19F pulses through Amp 1
Channel 1 (high band), even when running 2D 1H/19F experiments.

In this case, there is a difference in PTS channels and amplifier channels. PTS will
use channel 1 and 2, the amplifier will only use channel 1 (high band).



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2586): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2586__;!!PvDODwlR4mBZyAb0!QfDRPAqaNF-h_rzbIDUTtVqINmM4Wy=
bxoRBIKVkM-XSKkHj0D1IRZGNkF1Sr9qtFBjxS2tA2JOHjcim-qr4pd34y-_RH$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/114278283=
/7559972__;!!PvDODwlR4mBZyAb0!QfDRPAqaNF-h_rzbIDUTtVqINmM4WybxoRBIKVkM-XSKk=
Hj0D1IRZGNkF1Sr9qtFBjxS2tA2JOHjcim-qr4pdylN8_m1$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Mon Jul 28 2025 - 14:33:27 MST