Re: [AMMRL] VT Problem _at_ Bruker 400 Avance II HD Nanobay

From: Andrew Fowler via groups.io <afowler13=gmail.com_at_groups.io>
Date: Fri, 25 Jul 2025 09:55:31 -0400

One quick thought - the filename of the attachment suggests you have a BSVT
as opposed to the older BVT. Presumably you then also have a VTA-TC2T on
the floor near the probe. The two TC-2T cables are identical but have
different circuits and it's not unheard of for one to go bad. I would try
swapping the channel 1 cable at the probe for the channel 2 cable, then
quit and restart edte and see if it behaves. Having said that, I don't
think I've seen a channel fail with such a large overtemp error but it's
worth trying.

Best,
Andrew

On Fri, Jul 25, 2025 at 9:37 AM Dr. Birgit Claasen via groups.io wrote:

> Dear all,
>
> we have currently a problem with the VT unit of our Bruker 400 Avance II
> HD Nanobay NMR spectrometer. Enclosed you can find a brief summary with
> some figures documenting the problem.
>
> The heater cannot be switched on in the EDTE window (error message:
> ‘Parameter VTC_CMD_STAT parameter not writable’). Channel 1 reports:
> ‘Regulator inoperative’. The probe gas is also off and cannot be
> switched on. There appears to be no connection to the probe.
>
> The log file history of the EDTE window shows that there was recently an
> overheating in the safety sensor: ‘1561.4 K is above limit of 673.0K on
> Ch1.’
>
> We then performed a VT selftest using the BSMS Service Tool. It says
> ‘Active Ch1 heater TC_2T inoperative. Load resistance too high: 722.32
> ohms.’
>
> We suspect that the heater in the probe or the TC_T2 controller is
> defective. To test this, we would like to measure the heating plug. What
> values can be expected here? We have a 7-pin plug .
>
> What else could we try? Does anyone have any ideas?
>
> Best wishes
>
> Birgit Claasen
>
> --
> Dr. Birgit Claasen
> Analytikabteilung - Technische Leitung
> Institut für Organische Chemie
> Universität Stuttgart
> Pfaffenwaldring 55
> 70569 Stuttgart
> Tel: +49 711 68564379
>
>


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2563): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2563__;!!PvDODwlR4mBZyAb0!RJq_hKockN4MXvJn2W-O9ZwZ64Ht29=
Ak1r5RHdlTp_y4btsDU0CisYaveZYOS93XdbmDasFMNXiPAvSvpzMIY1vYwd07$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/114337661=
/7559972__;!!PvDODwlR4mBZyAb0!RJq_hKockN4MXvJn2W-O9ZwZ64Ht29Ak1r5RHdlTp_y4b=
tsDU0CisYaveZYOS93XdbmDasFMNXiPAvSvpzMIY14J8Wrd$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Fri Jul 25 2025 - 07:22:12 MST

This archive was generated by hypermail 2.4.0 : Sat Aug 02 2025 - 14:13:11 MST