Re: [AMMRL] 19F and 1H single channel HOESY (Varian)

From: Shawn Wagner via groups.io <shawn.wagner=usc.edu_at_groups.io>
Date: Wed, 23 Jul 2025 18:06:53 +0000

Thanks Bert, David and Baa for the suggestions. I have a 400MR, but it is c=
urrently fit with a OneProbe that is not dual tuned for 1H-19F. I will lik=
ely tie the Channel 1 and 2 (on the VNMRS) together with a combiner/splitte=
r. This seems to be the lowest energy solution since it requires just a sma=
ll hardware change. It's not a core magnet so I don't want to make too many=
 changes. They will be happy if it works since they do a lot of fluorine ch=
emistry.

________________________________
From: main_at_ammrl.groups.io <main_at_ammrl.groups.io> on behalf of Spin-Doc <be=
rt.heise_at_spin-doc.net>
Sent: Tuesday, July 22, 2025 2:29 PM
To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
Subject: Re: [AMMRL] 19F and 1H single channel HOESY (Varian)

This Message Is From an External Sender
USC detected that this message came from outside of USC.
Report Suspicious<https://us-phishalarm-ewt.proofpoint.com/EWT/v1/LIr3w8kk_=
Xxm!MtbpLGS61aAkeGp5oDmUaoh-XWY7V7ExfPSDmYie449rXyd2m_2KORupShKEgwych6CJVhj=
-McwJ8uJSIxq4WNXER0LJ2jtUGZgcYl5T$ >

Hi David,

you could try to get the sample used in fidlib/mollib ;-)
(a chemist told me I may not have gotten the molecule's name fully correct… ;-)

That said, one quick related note: While I never measured this, apparently
in DD2 consoles the short bridging cable which enables these kinds of experiments
H{F}, F{H}, FH-HOESY & FH-HETCOR on a single-highband channel actually may
introduce a little bit of extra noise (not sure on which kinds of experiments -
maybe someone here did this kind of study?). So some users (or even Varian
engineers from what I heard) did remove this cable because of that. It is,
however, mandatory to have if you want to run these one-channel H-F experiments.
So in case you try to run them on a DD2 and they don't work - check whether the
bridge cable is actually present. Should go from Ch2 XMTR Aux Out to Ch1 XMTR Aux In.

Cheers
Bert


Am 22.07.2025 um 16:32 schrieb David Rovnyak via groups.io

this is a well-timed threat, to piggy-back on this, private replies fine to
spare people's inboxes, anybody recommend a favorite setup sample for the
HF HOESY? -David

--
---------------------------------------------------------------------------
David Rovnyak, Professor of Chemistry, Bucknell Univ., Lewisburg, PA 17837
BlueSky : NMR_Guy<https://bsky.app/profile/nmrguy.bsky.social>
Research: http://www.facstaff.bucknell.edu/drovnyak/>
Bucknell Chemistry: http://www.departments.bucknell.edu/chemistry/>
Pascal: "I would like to have written a shorter letter, but I didn't have the time"
---------------------------------------------------------------------------

Magnetic resonance equipment, service, training & more
Selling & buying used NMR equipment (all vendors)
Dr. rer. nat. Bert Heise
Owner, Applications Scientist

Spin-Doc Bert Heise
Kiefernweg 13
58239 Schwerte
Germany
Mobile: +49 171 5135315
Email: bert.heise_at_spin-doc.net<mailto:bert.heise_at_spin-doc.net>
Web: http://www.spin-doc.net

Representing:
Stelar Fast Field Cycling NMR Relaxometers (http://www.stelar.it)
MestreLab NMR Software (http://www.mestrelab.com)
Barthel HF Amplifiers (http://www.barthel-hf.de)




-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2536): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2536__;!!PvDODwlR4mBZyAb0!VjxtZTdkY6hn3pc04wupzro9EsttIS=
uzr_UIPgTmUvs5kigXXoddAdYB1R7z0_r80urXN3mLDTK8SIhwJMA4boW3vsik$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/114278283=
/7559972__;!!PvDODwlR4mBZyAb0!VjxtZTdkY6hn3pc04wupzro9EsttISuzr_UIPgTmUvs5k=
igXXoddAdYB1R7z0_r80urXN3mLDTK8SIhwJMA4bvq5P2m7$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Jul 23 2025 - 14:06:49 MST

This archive was generated by hypermail 2.4.0 : Sat Aug 02 2025 - 14:13:11 MST