From: Spin-Doc <bert.heise_at_spin-doc.net>
Date: Tue, 22 Jul 2025 23:29:33 +0200

Hi David,

you could try to get the sample used in fidlib/mollib ;-)
(a chemist told me I may not have gotten the molecule's name fully correct… ;-)

That said, one quick related note: While I never measured this, apparently
in DD2 consoles the short bridging cable which enables these kinds of
experiments H{F}, F{H}, FH-HOESY & FH-HETCOR on a single-highband channel
actually may introduce a little bit of extra noise (not sure on which kinds
of experiments - maybe someone here did this kind of study?). So some users
(or even Varian engineers from what I heard) did remove this cable because
of that. It is, however, mandatory to have if you want to run these one-channel
H-F experiments. So in case you try to run them on a DD2 and they don't work -
check whether the bridge cable is actually present. Should go from Ch2
XMTR Aux Out to Ch1 XMTR Aux In.

Cheers
Bert


> Am 22.07.2025 um 16:32 schrieb David Rovnyak via groups.io:
>
> this is a well-timed threat, to piggy-back on this, private replies fine
> to spare people's inboxes, anybody recommend a favorite setup sample for
> the HF HOESY? -David
>
> --
> -------------------------------------------------------------------------
> David Rovnyak, Professor of Chemistry, Bucknell Univ., Lewisburg, PA 17837
> BlueSky : NMR_Guy <https://bsky.app/profile/nmrguy.bsky.social>
> Research: http://www.facstaff.bucknell.edu/drovnyak/
> Bucknell Chemistry: http://www.departments.bucknell.edu/chemistry/

>Pascal: "I would like to have written a shorter letter, but I didn't have the time"
> -------------------------------------------------------------------------
>


Magnetic resonance equipment, service, training & more
Selling & buying used NMR equipment (all vendors)
Dr. rer. nat. Bert Heise
Owner, Applications Scientist

Spin-Doc Bert Heise
Kiefernweg 13
58239 Schwerte
Germany
Mobile: +49 171 5135315
Email: bert.heise_at_spin-doc.net
Web: http://www.spin-doc.net

Representing:
Stelar Fast Field Cycling NMR Relaxometers (http://www.stelar.it
MestreLab NMR Software (http://www.mestrelab.com
Barthel HF Amplifiers (http://www.barthel-hf.de



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2528): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2528__;!!PvDODwlR4mBZyAb0!RQLS2MluT9YsLBIcbtI5BNEmKAQq12=
DWMmYhYULcHTOqskeQVWDCkG8HsCsGgVeqg_fsu4cetjsSFQPacz2wyA$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/114278283=
/7559972__;!!PvDODwlR4mBZyAb0!RQLS2MluT9YsLBIcbtI5BNEmKAQq12DWMmYhYULcHTOqs=
keQVWDCkG8HsCsGgVeqg_fsu4cetjsSFQOMRKo6EA$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-



Received on Tue Jul 22 2025 - 14:47:21 MST

This archive was generated by hypermail 2.4.0 : Sat Aug 02 2025 - 14:13:11 MST