Re: [AMMRL] Problem with BACS sample changer

From: Gopal Subramaniam via groups.io <gopal.subramaniam=qc.cuny.edu_at_groups.io>
Date: Tue, 22 Jul 2025 02:20:35 +0000

I wonder the problem is likely the connecting cable and next possibility
circuit board on the sample changer.

Gopal Subramaniam
Dept of Chem and Biochem
Queens College CUNY
Flushing, NY 11367
________________________________
> From: main_at_ammrl.groups.io on behalf of Philipp Hegemann via groups.io
> Sent: Friday, July 18, 2025 3:20 AM
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: [AMMRL] Problem with BACS sample changer

Dear members of the AMMRL community,

we have been experiencing a recurring issue with our BACS sample changer
connected to an Avance II 300 MHz spectrometer.

The error message “downwards motion failed” appears regularly, even though
the vertical cylinder moves down correctly.

In a first try, we replaced the entire vertical cylinder with another unit,
but unfortunately without success. The error continued to occur.

In a second attempt, we replaced the entire BACS sample changer with another
unit that had become available after decommissioning an older spectrometer.
This BACS unit had previously operated reliably. However, the same error
(“downwards motion failed”) occurred again, which led us to revert back
to the original BACS. The error still persists.

We also reconnected the RemoteControl port (and exchanged the cable) on the
BSMS to a different socket and updated the corresponding setting in “cf”.
The sample changer can be addressed correctly via IconNMR and the “rstest”
tool but the error continued to occur

In addition, we regularly receive the error message “Sample Detect at Magnet
Failed” when attempting to place the sample on the lift.

We would greatly appreciate any suggestions regarding the potential cause of
the issue or further diagnostic steps.

Greetings

Philipp Hegemann



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2521): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2521__;!!PvDODwlR4mBZyAb0!XnDIJovdDML6rR6rJ-ve01Xs6Idvqs=
9JCIZuNIDm0KWhWHid1paJU-mLF1E53cQYiqyaF6geMETpINgkb5oIuMNLX1yb$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/114278278=
/7559972__;!!PvDODwlR4mBZyAb0!XnDIJovdDML6rR6rJ-ve01Xs6Idvqs9JCIZuNIDm0KWhW=
Hid1paJU-mLF1E53cQYiqyaF6geMETpINgkb5oIuPJu3Zk5$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Mon Jul 21 2025 - 20:24:52 MST

This archive was generated by hypermail 2.4.0 : Sat Aug 02 2025 - 14:13:11 MST