Re: [AMMRL] GAB_Error in Avance III BSMS #Bruker #Hardware

From: Craig Grimmer <craig.grimmer_at_gmail.com>
Date: Fri, 5 Jul 2024 11:35:03 +0200

Hello Anne

It seems like the GAB/2 unit is defective. If you have another instrument
with a GAB/2, you could try swapping out the suspect GAB/2 with a
known-to-be-working GAB/2 from another spectrometer. If you have a
different probe for the instrument, you could test for a probe fault.
Ditto for the gradient cable.

Regards,
Craig.




On Wed, 3 Jul 2024 at 14:59, Michael Groves via groups.io <mgrovesnmr=
yahoo.com_at_groups.io> wrote:

> Hi all,
>
> Just my $0.02--the GAB/2 isn't smart enough to check the gradient circuit
> in the probe--if it is going into an error state when you're not pulsing it
> sounds like the GAB is bad. You can check the gradient in the probe--if
> it's a normal room temperature probe you can check the resistance between
> pins A-B and it should be around 1.5 ohms. And check the resistance to
> ground--either from A or B to the center pin of the connector or to the
> shell around the connector, and if it's not extremely high (it might take a
> few seconds to ramp up to its final value) then there could be an issue
> with the gradient shorting to ground. I suspect you'll find that's not the
> case.
>
> I suppose I could imagine some pathological case where something was
> shorting to the probe and creating a voltage somewhere there shouldn't be
> one--you could always try resetting the GAB with the gradient cable
> disconnected and confirming that it behaves the same.
>
> Cheers,
> Mike
>
> On Wednesday, July 3, 2024 at 04:17:08 AM MDT, Ulrich <
> ulrich.haunz_at_uni-konstanz.de> wrote:
>
>
> Hello,
>
> I had also troubles with the GAB board.
>
> ha -> bsms -> gradients -> Gradient Amplifier Control
>
> there on the control panel - are some buttons. There you can set the
> Gradient unit ON or OFF.
> On the next page - you can do some diagnostic tests and Resets.
>
> This helped me a lot. Or you can update the firmware for the GAB-Boad.
> Via the BSMS Service Tools you can look, if the whole BSMS rack is OK or
> something is old.
> Also you can easily access there the Firmware update routine for the
> GAB-Board, too.
>
> gab2_fpga_00-013-02.bit is my actual firmware for the GAB/2 board.
>
> The worst case is - the gradient is broken in the probe. There is a big
> round plug outside of the probe.
> This is the gradient plug. There you can easily measure the contacts
> through the gradient. There should be a stable resistive value.
> If it is broker - there are all contacts at 0 Ohm.
>
> Yours sincerely,
>
> Ulrich Haunz
>
> Am Dienstag, Juli 02, 2024 21:02 CEST, schrieb "Bob Berno" :
>
> > Hello Anne,
> >
> > I have seen this and similar errors a few times in the past. On a few
> > occasions it was a problem with the gradient circuit in the probe, and
> > once it was due to a failure in the GAB board.
> >
> > From your symptoms, I think you have a problem with the GAB board.
> > Do you know if the gradients work in any other experiments?
> > What happens when you try to run an experiment with uses pulsed field
> > gradients (PFGs) for coherence selection (most 2D experiments will be
> > setup to use PFGs by default)? Do you see the lock level go down and
> > then recover as the experiment runs?
> >
> > Good luck,
> > Bob.
> >
> > On 2024-07-02 2:06 p.m., Anne Carroll via groups.io wrote:
> > >
> > >
> > > Hello,
> > > Please let me know if you have encountered this issue and how your
> > > resolved it.
> > >
> > > In TopSpin 3.6.2, the issue manifests this error "TopshimRawCheck -
> > > the Z gradient does not work" whenever we run gradient shimming.
> > > Non-gradient shimming (topshim tune) runs fine. There is nothing wrong
> > > with the probe.
> > > When I reboot the Gradient Amplifier Board GAB/2 from the Service
> > > Page, I hear a click and the Ready light on the board turns green. If
> > > I Refresh the web service page while the light is green, the Board
> > > State claims 'Operation.' Within a few seconds, I hear another click
> > > and the Ready light goes out. Now the Refresh shows 'GAB_Error.'
> > >
> > > Thank you,
> > > Anne Carroll
> > >
> > > --
> > > Anne M. Carroll, Ph.D. (she/her) NMR/MS Instrumentation Specialist
> > > Department of Chemistry and Biochemistry Miami University Oxford, Ohio
> > > USA513-529-3163 Office 678-637-6739 Cell carrol50_at_miamioh.edu
> > >
> >
> > --
> > Dr. Bob Berno
> > Manager, NMR Facility
> > McMaster University, ABB-156
> > 1280 Main St. West
> > Hamilton, ON L8S 4M1
> > T: 905-525-9140 x24158 or x24686
> >
> -------------------------------------------------------------------------
> > This email contains CONFIDENTIAL information intended for the specific recipient(s).
> > If you have an accommodation need for a planned meeting, please e-mail
> > me directly and
> > I will do my best to make appropriate arrangements.
> > Should you require any materials sent via this e-mail address in an alternate/accessible
> > format, please let me know.
> >
> > McMaster University sits on the traditional Territories of the Mississauga and
> > Haudenosaunee Nations, and within the lands protected by the "Dish With One
> > Spoon" wampum agreement.
> >
> --
> =========================
>
> Ulrich Haunz, Dipl.-Ing. (FH)
> NMR-Spektroskopie
>
> Raum L 710, L 807
> NMR-Center M520, M521, M527
> Postfach M 715
> Fachbereich Chemie
> Universitaet Konstanz
>
> 78457 Konstanz
>
> Tel.: +49 (0)7531 88-4261 -> M521 Bruker Avance NEO 800
> Tel.: +49 (0)7531 88-4422 -> L807 JEOL Lambda 400
> Tel.: +49 (0)7531 88-4955 -> L710 Bruker AVIII HD 400
> Tel.: +49 (0)7531 88-4956 -> M520a Bruker AVIII 600
> Tel.: +49 (0)7531 88-4957 -> M527 Bruker AVIII 400 Automation
> Tel.: +49 (0)7531 88-4958 -> M520 Bruker AVIII 400 WB
> Tel.: +49 (0)170 9153213
> FAX : +49 (0)7531 88-3898
> =========================
>


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1394): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1394__;!!PvDODwlR4mBZyAb0!VWXmz88d2tC1Ci6m52SttF5SBy0k08=
2iV3WdQKy_gMON2nppvTGClluoxAnN4y1kIZpZbmD8U9SR5QpKXEHpQw$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/107005267=
/7559972__;!!PvDODwlR4mBZyAb0!VWXmz88d2tC1Ci6m52SttF5SBy0k082iV3WdQKy_gMON2=
nppvTGClluoxAnN4y1kIZpZbmD8U9SR5QrVPnxSrQ$
Mute #bruker:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mut=
ehashtag/bruker__;!!PvDODwlR4mBZyAb0!VWXmz88d2tC1Ci6m52SttF5SBy0k082iV3WdQK=
y_gMON2nppvTGClluoxAnN4y1kIZpZbmD8U9SR5QqrqfuaiA$
Mute #hardware:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/m=
utehashtag/hardware__;!!PvDODwlR4mBZyAb0!VWXmz88d2tC1Ci6m52SttF5SBy0k082iV3=
WdQKy_gMON2nppvTGClluoxAnN4y1kIZpZbmD8U9SR5QokBezyvA$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Fri Jul 05 2024 - 02:35:21 MST

This archive was generated by hypermail 2.4.0 : Sat Aug 03 2024 - 15:02:58 MST