Re: [AMMRL] GAB_Error in Avance III BSMS #Bruker #Hardware

From: Anne Carroll <carrol50_at_miamioh.edu>
Date: Tue, 09 Jul 2024 11:57:41 -0700

Update.

I have now tested the faulty GAB/2 in our other Avance III console and got
the same failure conditions, so the gradient board is definitively the problem.

I will appreciate any recommendations for sellers of second-hand boards.

Thank you,
Anne
--