Re: [AMMRL] Water or Air cooled Cryoplatform?

From: Parnham, Stu <stitch_at_email.unc.edu>
Date: Thu, 29 Feb 2024 16:46:11 +0000

Hi Justin,

I currently have water cooled but in the past I've had indoor and outdoor air
cooled. I love the water cooled and have had none of the issues I had with
the air cooled, no cleaning of the fins every 3 months. The only thing I
would say is to make sure that your chilled water temperatures do not fluctuate
too much and that there is no particulates floating in the system. The
facility guys usually have good numbers to hand for the chilled water and
you can always run them past Walter Osgood at Bruker for approval.

Stu

-------------------
Stuart Parnham Ph.D.
Biomolecular NMR
Marsico Hall, SB205A
University of North Carolina
Phone:(919)-966-6184

> From: main_at_ammrl.groups.io On Behalf Of Justin Pontius via groups.io
> Sent: Thursday, February 29, 2024 11:04 AM
> To: main_at_ammrl.groups.io
> Subject: [AMMRL] Water or Air cooled Cryoplatform?

Hi Everyone,

We're in the planning stages of upgrading our aging AV1-800 to something more
modern. Currently we have a 1st gen cryoplatform with an air cooled compressor
outside on a balcony. Maintenance is a small pain with all the tree leaves,
cotton, and dirt. If you had the choice would you go with another air cooled
compressor or water cooled? We don't have chilled water in the lab yet but
it's about 50 feet away in a service closet so it wouldn't be too bad to
bring it in. Thoughts?

--Justin

Justin Pontius
NMR Facility Engineer
Magnetic Resonance Research Center
University of Notre Dame

574-631-7601 (direct line)

145A Nieuwland Science Hall
Notre Dame, IN 46556



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1035): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1035__;!!PvDODwlR4mBZyAb0!QFy0cxgMLCJ1pbuofZ3Hbycm3uWT-r=
EI3MVg54ejyZ-ZVJuG7EZwhHTJKmyvhPRB4yGkk1MlU4r4Ea_N18c$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/104646404=
/7559972__;!!PvDODwlR4mBZyAb0!QFy0cxgMLCJ1pbuofZ3Hbycm3uWT-rEI3MVg54ejyZ-ZV=
JuG7EZwhHTJKmyvhPRB4yGkk1MlU4r4Nv8fli8$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Thu Feb 29 2024 - 08:46:17 MST

This archive was generated by hypermail 2.4.0 : Thu Feb 29 2024 - 14:59:21 MST