Re: [AMMRL] Water or Air cooled Cryoplatform?

From: Alex Paterson via groups.io <alpaterson=wisc.edu_at_groups.io>
Date: Thu, 29 Feb 2024 16:58:06 +0000

Hi Justin,

We run about 13 water-cooled helium compressors (mostly Sumitomo F-50 and F-70
units) and would not in any circumstances change to air. The technical manuals
for the compressors have good information for allowable chilled water temperature
and flow rates, and they’re fairly flexible for those parameters.
I third the caution about water quality, though. We have a closed process chilled
water loop that rejects the heat to campus chilled water, which avoids most water
quality issues. If campus chilled water doesn’t meet the quality spec,
you might want to look into that as an option.

Best,
Alex

Alexander Paterson, PhD
Scientist II | Department of Biochemistry
The National Magnetic Resonance Facility at Madison (NMRFAM)
University of Wisconsin-Madison
433 Babcock Dr, Room B160C | Madison, WI 53706
TEL: (608) 263-0017


From: main_at_ammrl.groups.io <main_at_ammrl.groups.io> On Behalf Of Parnham, Stu
Sent: Thursday, February 29, 2024 10:46
To: main_at_ammrl.groups.io
Subject: Re: [AMMRL] Water or Air cooled Cryoplatform?

Hi Justin,


I currently have water cooled but in the past I’ve had indoor and
outdoor air cooled. I love the water cooled and have had none of the issues
I had with the air cooled, no cleaning of the fins every 3 months. The only
thing I would say is to make sure that your chilled water temperatures do not
fluctuate too much and that there is no particulates floating in the system.
The facility guys usually have good numbers to hand for the chilled water
and you can always run them past Walter Osgood at Bruker for approval.

Stu

-------------------
Stuart Parnham Ph.D.
Biomolecular NMR
Marsico Hall, SB205A
University of North Carolina
Phone:(919)-966-6184

> From: main_at_ammrl.groups.io On Behalf Of Justin Pontius via groups.io
> Sent: Thursday, February 29, 2024 11:04 AM
> To: main_at_ammrl.groups.io<mailto:main_at_ammrl.groups.io>
> Subject: [AMMRL] Water or Air cooled Cryoplatform?


Hi Everyone,

We're in the planning stages of upgrading our aging AV1-800 to something more
modern. Currently we have a 1st gen cryoplatform with an air cooled compressor
outside on a balcony. Maintenance is a small pain with all the tree leaves,
cotton, and dirt. If you had the choice would you go with another air cooled
compressor or water cooled? We don't have chilled water in the lab yet but it's
about 50 feet away in a service closet so it wouldn't be too bad to bring it in.
Thoughts?

--Justin

Justin Pontius
NMR Facility Engineer
Magnetic Resonance Research Center
University of Notre Dame
nmr.nd.edu

574-631-7601 (direct line)

145A Nieuwland Science Hall
Notre Dame, IN 46556



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1037): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1037__;!!PvDODwlR4mBZyAb0!Wq3Zc34Xyc1JtJx6fOgz3VhXDXM1Oh=
2pbUn6NbPyIlOdz3uEHQxtvWHET50KLav9I_H3pPw5GKF1nsVev-dDj1rzmkQD$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/104646404=
/7559972__;!!PvDODwlR4mBZyAb0!Wq3Zc34Xyc1JtJx6fOgz3VhXDXM1Oh2pbUn6NbPyIlOdz=
3uEHQxtvWHET50KLav9I_H3pPw5GKF1nsVev-dDj8HDfe2b$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Thu Feb 29 2024 - 08:58:14 MST

This archive was generated by hypermail 2.4.0 : Thu Feb 29 2024 - 14:59:21 MST