Re: [AMMRL] Air compressor dew point -40C or -70C for NMR with BCU1 and BCU2

From: Nick Fitzkee via groups.io <nfitzkee=chemistry.msstate.edu_at_groups.io>
Date: Tue, 27 Feb 2024 14:49:22 +0000

You can also purchase equipment to lower the dewpoint from a normal compressor.
We do this in our facility. Ballston is main manufacturer. I’ve attached
a picture of our setup below. If you attach an N2 separator in line before your
console, it’s essentially the same as having a building nitrogen source.


--
Nicholas Fitzkee
Professor of Chemistry
Director, MSU NMR Facility
Mississippi State University
nfitzkee_at_chemistry.msstate.edu
http://fitzkee.chemistry.msstate.edu/
(662) 325-1288 | Twitter: _at_FitzkeeLab _at_chem_msstate

NMR and Mass Spec Facility Manager Positions (Spring 2024)
https://www.chemistry.msstate.edu/about/employment*managers

> From: main_at_ammrl.groups.io <main_at_ammrl.groups.io> On Behalf Of Thomas Alber
> Sent: Friday, February 23, 2024 12:18 PM
> To: main_at_ammrl.groups.io
> Subject: Re: [AMMRL] Air compressor dew point -40C or -70C for NMR with BCU1 and BCU2

Hello!

Many people that I know run their NMR off the nitrogen boiloff from a lN2 tank.
If that is a possibility for you, this will provide nitrogen with the required low dewpoint.

Thomas


> From: main_at_ammrl.groups.io> On Behalf Of Xia, Youlin via groups.io
> Sent: Friday, February 23, 2024 10:11 AM
> To: main_at_ammrl.groups.io<mailto:main_at_ammrl.groups.io>
> Subject: [AMMRL] Air compressor dew point -40C or -70C for NMR with BCU1 and BCU2

Dear all,

We are considering purchasing an air compressor working for both 400 MHz (room
temp probe with BCU1; ordered) and 700 MHz (cryoprobe with BCU2; currently working
with a small air compressor temporarily). From Bruker’s manual, BCU1 requires
a dew point of -50C for the compressed gas and BCU2 requires -80C. Now the situation
we are facing is the air compressor with dew point of -70C is much expensive and with
several months lead time. The one with the dew point of -40C is available and cheaper.
We hope to get the 400 MHz installed as early as possible and can’t wait the
one with -70C ready.

We really appreciate you can provide us your experience and advice on choosing the
right air compressor.

1, Does the air compressor with dew point of -40C work well for your machines with BCU2?
2, Is a dryer a must have if you use the air compressor with dew point of -40C? If yes,
could you please provide the details (brand, model, cost, reliability, etc.)?
3, Do you think it must use the air compressor with dew point of -70C for NMR machines
with BCU2 (probable we only found one type with -70C)?
4, Could you please recommend the air compressor with dew point lower than -70C if we must to have?
5, Any other advice for the compressed air supply with such required quality for our case
(equipped with BCU1 and BCU2)?

Really appreciate your information!
Thanks,
Fahu
Fahu.he_at_slu.edu<mailto:Fahu.he_at_slu.edu>


________________________________


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1020): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1020__;!!PvDODwlR4mBZyAb0!VchjUfqUjwRUtZvPf-jNuQd6fCo0lz=
JzRO1h9jcyzWbsUyEOtMCZ-Zur9UFIYGsnAnDeDSNZxX67_qQHxRKdcblVE6jO$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/104533230=
/7559972__;!!PvDODwlR4mBZyAb0!VchjUfqUjwRUtZvPf-jNuQd6fCo0lzJzRO1h9jcyzWbsU=
yEOtMCZ-Zur9UFIYGsnAnDeDSNZxX67_qQHxRKdcfMMdrGk$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-






image003.jpg
(image/jpeg attachment: image003.jpg)

image004.jpg
(image/jpeg attachment: image004.jpg)

Received on Tue Feb 27 2024 - 06:52:34 MST

This archive was generated by hypermail 2.4.0 : Thu Feb 29 2024 - 14:59:21 MST