Dear Lisa Tuttle
If you have another NMR spectrometer available nearby, you could use its 10MHz
output of REF board. I have encountered this problem for Bruker Avance 400
spectrometer five years ago, I use the 10 MHz output from Bruker Avance 600 NMR
spectrometer nearby. This works very well except Avance 400 have to stop work
when Avance 600 have problem. The problem have been finally so lved after I
upgrade the consoles for two machines.
Best wishes
Junfeng Xiang
ICCAS, Beijing China
-----原始邮件-----
发件人:Spin-Doc <bert.heise_at_spin-doc.net>
发送时间:2026-08-21 15:47:16 (星期五)
收件人: main_at_ammrl.groups.io
抄送:
主题: Re: [AMMRL] AQS only loads correctly with external 10 MHz signal
Hi,
one worthwhile check is the power supply board at the back which should provide
the (usually 15 V I think) correct voltage for the synthesizer's oscillator
crystal "oven". I have seen it before that a synthesizer may produce the wrong
frequency if the voltage is no longer adequate.
Cheers
Bert
Am 20.08.2026 um 22:57 schrieb Lisa Tuttle via groups.io:
We had the AQS fan tray fail on our 500 MHz Avance III earlier this summer.
Some of the SGU boards in the AQS reported an overheat error. I've replaced
the fan rack, but since then, the AQS gives flashing errors across all boards
(RXAD, REF600, and 4xSGU2) and the boards are not recognized in the AQS web
portal or through 'cf' in Topspin. 'cf' hangs at an HPPR step.
We determined that the REF600 board was outputting 10.92 MHz instead of the
necessary 10 MHz. With an external signal generator connected to the REF600
to supply the 10 MHz signal, the output signal is the desired 10 MHz, the
AQS boards will load correctly, and we can run experiments.
So we replaced the REF600 board - but the same thing still happens. We also
tried different power supply boards, no change. This 10.9 MHz output is
reproducible with only the REF600 board connected on the AQS rack (all other
boards pulled out, and cables disconnected).
I think Bruker support is running out of things for me to test, so I turn to
you all.
Any ideas?
Thanks!
Lisa Tuttle
Klevit Lab, Research Scientist IV
University of Washington, Seattle
Magnetic resonance equipment, service, training & more
Selling & buying used NMR equipment (all vendors)
Dr. rer. nat. Bert Heise
Owner, Applications Scientist
Spin-Doc Bert Heise
Kiefernweg 13
58239 Schwerte
Germany
Mobile: +49 171 5135315
Email: bert.heise_at_spin-doc.net
Web: http://www.spin-doc.net
Representing:
Stelar Fast Field Cycling NMR Relaxometers (
http://www.stelar.it )
MestreLab NMR Software (
http://www.mestrelab.com )
Barthel HF Amplifiers (
http://www.barthel-hf.de )
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3710): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3710__;!!PvDODwlR4mBZyAb0!ST2eUPS9AWEwLgs8AxLMnNyBHo6iVl=
ZXOmt6_aix_4pF3T1oHhm51399riGZp9Uf5l-4EG0SlcEXybTNaw$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120853723=
/7559972__;!!PvDODwlR4mBZyAb0!ST2eUPS9AWEwLgs8AxLMnNyBHo6iVlZXOmt6_aix_4pF3=
T1oHhm51399riGZp9Uf5l-4EG0SlcGdQ6FrwQ$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Sun Aug 23 2026 - 18:37:16 MST