Hi,
one worthwhile check is the power supply board at the back which should provide
the (usually 15 V I think) correct voltage for the synthesizer's oscillator
crystal "oven". I have seen it before that a synthesizer may produce the
wrong frequency if the voltage is no longer adequate.
Cheers
Bert
> Am 20.08.2026 um 22:57 schrieb Lisa Tuttle via groups.io:
>
> We had the AQS fan tray fail on our 500 MHz Avance III earlier this summer.
> Some of the SGU boards in the AQS reported an overheat error. I've replaced
> the fan rack, but since then, the AQS gives flashing errors across all boards
> (RXAD, REF600, and 4xSGU2) and the boards are not recognized in the AQS web
> portal or through 'cf' in Topspin. 'cf' hangs at an HPPR step.
>
> We determined that the REF600 board was outputting 10.92 MHz instead of the
> necessary 10 MHz. With an external signal generator connected to the REF 600
> to supply the 10 MHz signal, the output signal is the desired 10 MHz, the AQS
> boards will load correctly, and we can run experiments.
>
> So we replaced the REF600 board - but the same thing still happens. We also
> tried different power supply boards, no change. This 10.9 MHz output is
> reproducible with only the REF600 board connected on the AQS rack (all other
> boards pulled out, and cables disconnected).
>
> I think Bruker support is running out of things for me to test, so I turn to
> you all.
>
> Any ideas?
>
> Thanks!
>
> Lisa Tuttle
> Klevit Lab, Research Scientist IV
> University of Washington, Seattle
>
Magnetic resonance equipment, service, training & more
Selling & buying used NMR equipment (all vendors)
Dr. rer. nat. Bert Heise
Owner, Applications Scientist
Spin-Doc Bert Heise
Kiefernweg 13
58239 Schwerte
Germany
Mobile: +49 171 5135315
Email: bert.heise_at_spin-doc.net
Web: http://www.spin-doc.net
Representing:
Stelar Fast Field Cycling NMR Relaxometers (
http://www.stelar.it)
MestreLab NMR Software (
http://www.mestrelab.com)
Barthel HF Amplifiers (
http://www.barthel-hf.de)
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3704): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3704__;!!PvDODwlR4mBZyAb0!VZybqDwd-V87-F811zKuma6Y3L4DKL=
-8UTcU1afU0jdJQR1Zd_s9zG_Bw3Jk5KQFoBkUP6JN_1utlXWm4c0Alg$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120853723=
/7559972__;!!PvDODwlR4mBZyAb0!VZybqDwd-V87-F811zKuma6Y3L4DKL-8UTcU1afU0jdJQ=
R1Zd_s9zG_Bw3Jk5KQFoBkUP6JN_1utlXXbDNULlA$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Fri Aug 21 2026 - 03:03:23 MST