Hi TJ,
there should be no concern. It is the shim system that has to fit in the magnet.
And almost all magnets have a room temperature bore of 51 mm or more. Every
vendors shims will fit and once the shims are in all probes will fit as well.
The pages you included say 54mm. So no problem. The second important dimension
is the distance of the magnetic center from the bottom plate. Your pages say
320+/- 3mm. Our maximum distance is 332 mm so no problem either.
Best regards
Clemens
P.S. for correctness I want to mention that there used to be magnets with
44mm bore mostly at the 300 and 400 level. And I am aware of set of Oxford
500's with less than 51 mm bore.
On 8/21/2026 11:04 AM, Thusitha Jayasundera via groups.io wrote:
Hello all--has anyone updated their Varian/Agilent MR-400 magnet with a Bruker
or Jeol solid state probe and console? If you have, is the setup working well
for you? Any advice before we make this move?
I'm trying to ascertain for sure if this magnet can be used with the probes
from each vendor (i.e. will they fit).The pages with information on the magnet
itself are attached; our current setup is an AutoX broadband probe for regular,
liquid-state NMR.
Any insights would be much appreciated--have a good weekend!
TJ.
--
==========================
==========================
====================
Clemens Anklin Ph.D. Vice President
Bruker BioSpin NMR Applications & Training
15 Fortune Drive NEW Phone: 978-313-5NMR(5667)
Billerica MA 01821 bluesky: _at_canklin.bsky.social
web: http://www.bruker.com
e-mail: clemens.anklin_at_bruker.com
==========================
==========================
====================
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3709): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3709__;!!PvDODwlR4mBZyAb0!QTTIfT6rqHjVIyIkAADGVWZORVMZJj=
eEcdUMKOJrYh_iGHglLfNiVJG7ZWs3bP4_YFcsUMuUvDH8Ub8zR9_c1YkJK1vp$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120862906=
/7559972__;!!PvDODwlR4mBZyAb0!QTTIfT6rqHjVIyIkAADGVWZORVMZJjeEcdUMKOJrYh_iG=
HglLfNiVJG7ZWs3bP4_YFcsUMuUvDH8Ub8zR9_c1ZpJceNS$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Fri Aug 21 2026 - 11:45:49 MST