Hi Lisa,
It could be a problem with the back panel inside the AQS chassis where the
board are attached to.
The only way to find out is to swap the chassis itself.
Cheers,
Gerard.
Gerard J. A. Kroon
Manager NMR facility
The Scripps Research Institute
Core Facilities
Tel: +1 858-784-7485
Cell: +1 619-538-9818
Email: kroon_at_scripps.edu
________________________________
> From: main_at_ammrl.groups.io on behalf of Lisa Tuttle via groups.io
> Sent: Friday, August 21, 2026 8:50 AM
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: Re: [AMMRL] AQS only loads correctly with external 10 MHz signal
Thanks, we did try replacement AQS power supplies (PSM1 and PSM5) and the
behavior was unchanged.
-=-=-=-=-=-=-=-=-=-=-=-