Hi,
Searching Ebay for “MHz signal generator” (available for as little as $100-200)
I’d wager that using one of these is likely the cheapest option for a repair…
Might indeed be a bent pin on the backplane somewhere - you only swapped the fan tray at the beginning?
For a new one I guess, so all fans are now working, right?
If you check the 10 MHz output with a scope after switching on the console after leaving it
off for maybe one hour to cool down do you observe correct 10 MHz output at first, followed by
drifting away after a while? That might indicate still having some cooling issues…
Cheers
Bert
-> Sent via mobile <-
Spin-Doc
Magnetic resonance equipment, service, training & more
Dr. rer. nat. Bert Heise
Owner, applications scientist
Kiefernweg 13
58239 Schwerte
Germany
Mail: bert.heise_at_spin-doc.net
Phone: +49-171-5135315
Representing MestreLab software, Stelar relaxometers & Barthel HF amplifiers
Am 24.08.2026 um 03:37 schrieb 向俊锋 <jfxiang_at_iccas.ac.cn>:
Dear Lisa Tuttle
If you have another NMR spectrometer available nearby, you could use its 10 MHz output of REF board.
I have encountered this problem for Bruker Avance 400 spectrometer five years ago, I use the 10 MHz
output from Bruker Avance 600 NMR spectrometer nearby. This works very well except Avance 400 have to
stop work when Avance 600 have problem. The problem have been finally solved after I upgrade the
consoles for two machines.
Best wishes
Junfeng Xiang
ICCAS, Beijing China
_._,_._,_
Groups.io Links:
You receive all messages sent to this group.
View/Reply Onl=
ine (#3711) |
Reply to Group
| Reply to Sender
|
Mute This Topic=
a>
| New Topic
Your Subscription |
Contact Group Owner |
_._,_._,_
Received on Mon Aug 24 2026 - 03:12:01 MST
This archive was generated by hypermail 2.4.0
: Tue Sep 01 2026 - 13:53:27 MST