[AMMRL] PTS Synthesizer issue on Agilent DD2 #agilent #Hardware #Troubleshooting

From: Holmes, Daniel via groups.io <holmesd5=msu.edu_at_groups.io>
Date: Tue, 19 Aug 2025 06:38:18 -0700

We lost proton signal on our Agilent 500 MHz DD2 spectrometer, not even tune.
Traced the issue to the PTS output for high band. Scope from back of synthesizer
gives random low mV signal. Low band is good with excellent stability and power.
Ran testController script with no issues. My question is two-fold. Is this
definitely a synthesizer issue or could it be the RF controller? I plan to
swap the low band for high band controller to see if this moves the issue.
If it turns out to be the synthesizer, does anyone have one I could buy? I
have many spares for the Inova generation, but they do not use Ethernet
connections.


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2734): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2734__;!!PvDODwlR4mBZyAb0!UOSVtNBkOIkCktM9xglLBFG4WYE2Uc=
TBxMhcd0QF5IDv9Bou5fOZz2TtYwxusQAjteXgVkFanAKZwzI-aXvxR2iL8XbT$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/114781283=
/7559972__;!!PvDODwlR4mBZyAb0!UOSVtNBkOIkCktM9xglLBFG4WYE2UcTBxMhcd0QF5IDv9=
Bou5fOZz2TtYwxusQAjteXgVkFanAKZwzI-aXvxR-apfJ-K$
Mute #troubleshooting:https://urldefense.com/v3/__https://ammrl.groups.io/g=
/main/mutehashtag/troubleshooting__;!!PvDODwlR4mBZyAb0!UOSVtNBkOIkCktM9xglL=
BFG4WYE2UcTBxMhcd0QF5IDv9Bou5fOZz2TtYwxusQAjteXgVkFanAKZwzI-aXvxR0o1-X-P$
Mute #agilent:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mu=
tehashtag/agilent__;!!PvDODwlR4mBZyAb0!UOSVtNBkOIkCktM9xglLBFG4WYE2UcTBxMhc=
d0QF5IDv9Bou5fOZz2TtYwxusQAjteXgVkFanAKZwzI-aXvxRz0ycP3F$
Mute #hardware:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/m=
utehashtag/hardware__;!!PvDODwlR4mBZyAb0!UOSVtNBkOIkCktM9xglLBFG4WYE2UcTBxM=
hcd0QF5IDv9Bou5fOZz2TtYwxusQAjteXgVkFanAKZwzI-aXvxR5L03Zsk$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue Aug 19 2025 - 06:47:10 MST

This archive was generated by hypermail 2.4.0 : Mon Sep 01 2025 - 15:37:08 MST