Re: [AMMRL] PTS Synthesizer issue on Agilent DD2 #agilent #Hardware #Troubleshooting

From: Kim, Yong Wah via groups.io <Yong-Wah.Kim=utoledo.edu_at_groups.io>
Date: Tue, 19 Aug 2025 14:00:54 +0000

Hi Daniel,

Sounds like a problem I had with my Inova 600MHz. I have a 4 RF channel and
2 X PTS620 synthesizers. So, I swapped the synthesizers and the problem is
gone. The synthesizer output for H1 also improved dramatically. The H1
pulses shorten by 20% and the H1 tune pattern is back. Interestingly the
"defective" synthesizer works just fine at the lower frequencies. The
synthesizer probably handles the high, H1 frequency and the low / broad
band frequencies on different paths.

Regards

Yong-Wah Kim, PhD.





-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2735): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2735__;!!PvDODwlR4mBZyAb0!Sit-qEBkueo-DI1a3KGWyJE0Y3Mmu4=
9VO93dnBpJGSe5YJUHLBPcmcXVuztKN_7_OoJrx2DDk-b591S9AWi_qXXX-QWU$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/114781283=
/7559972__;!!PvDODwlR4mBZyAb0!Sit-qEBkueo-DI1a3KGWyJE0Y3Mmu49VO93dnBpJGSe5Y=
JUHLBPcmcXVuztKN_7_OoJrx2DDk-b591S9AWi_qZU08hED$
Mute #troubleshooting:https://urldefense.com/v3/__https://ammrl.groups.io/g=
/main/mutehashtag/troubleshooting__;!!PvDODwlR4mBZyAb0!Sit-qEBkueo-DI1a3KGW=
yJE0Y3Mmu49VO93dnBpJGSe5YJUHLBPcmcXVuztKN_7_OoJrx2DDk-b591S9AWi_qVd24ESZ$
Mute #agilent:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mu=
tehashtag/agilent__;!!PvDODwlR4mBZyAb0!Sit-qEBkueo-DI1a3KGWyJE0Y3Mmu49VO93d=
nBpJGSe5YJUHLBPcmcXVuztKN_7_OoJrx2DDk-b591S9AWi_qT9ltR9B$
Mute #hardware:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/m=
utehashtag/hardware__;!!PvDODwlR4mBZyAb0!Sit-qEBkueo-DI1a3KGWyJE0Y3Mmu49VO9=
3dnBpJGSe5YJUHLBPcmcXVuztKN_7_OoJrx2DDk-b591S9AWi_qVJgK_0n$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue Aug 19 2025 - 07:03:05 MST

This archive was generated by hypermail 2.4.0 : Mon Sep 01 2025 - 15:37:08 MST