Re: [AMMRL] ICON-NMR logins in Topspin 4.4.0

From: Bob Berno <bberno_at_mcmaster.ca>
Date: Thu, 15 Aug 2024 08:57:33 -0400

Hello David,

I think George has touched on a possible cause of your problem.
In addition to a pop-up window being reduced in size (which we have
seen), we have seen other instances where the pop-up window is behind
another window.  Sometimes you can sequentially minimize windows until
suddenly you see that pop-up window appear.  Other times, the pop-up
window minimizes along with the parent window (that is especially
annoying).  If the hidden pop-up window appears, you can click "OK" to
close the window. Often additional pop-up windows will appear.  Click
"OK" for all of the windows, and then eventually you will gain control
of IconNMR (and TopSpin) again.

I have not found a root cause of why pop-up windows appear in the
"background" behind other windows.  And I don't know why some systems
are more prone to it happening than others.
We haven't seen this for a while, so maybe there is a setting somewhere
that we selected or deselected, and the problem seems to have gone away.

The next time this happens, take a moment to look for the "hidden"
pop-up window(s).

Good luck,
Bob.

On 2024-08-14 7:01 p.m., Sukenick, George via groups.io wrote:
>
>
> I’ve seen something like this (TS 3.6.x; CentOS 7) where there is an
> “unseen” dialog box -
>
> for some reason, the dialog box, on occasion, is squished down to
> about the size (maybe smaller?) than a check box,
>
> hard to notice and then it seems like nothing responds because the
> software is expecting a response to the totally squished box.
>
> Of course, the solution is to expand the box and dismiss it or answer
> what it is asking for.
>
>   -------- -------- --------
>
> *George Sukenick, Ph.D.*
>
>        Associate Lab Member/Head
>
>        NMR Analytical Core Facility
>
> *       Memorial Sloan Kettering Cancer Center*
>
> *From: *main_at_ammrl.groups.io <main_at_ammrl.groups.io> on behalf of
> Vander Velde, David <davidv_at_caltech.edu>
> *Date: *Wednesday, August 14, 2024 at 6:04 PM
> *To: *main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> *Subject: *[EXTERNAL] [AMMRL] ICON-NMR logins in Topspin 4.4.0
>
> We've been having an issue where ICON-NMR will, seemingly at random,
> start nominally accepting logins, but the control buttons remain
> grayed out, and the box that says "Automation Locked" does not go away
> as it usually does with a login.
>
> We don't see this with TS 3.7.0, on another instrument with virtually
> the same user group. I have not been able to associate this with any
> particular user, or what they do, or don't do; or whether or not they
> have an ICON password.
>
> The one thing I'm sure is not involved is the operating system—I saw
> it first under CentOS 7, still see the same thing with Alma 9 on a new
> workstation.
>
> Is anyone else seeing this behavior? If so, has anyone found a
> workaround other than restarting the software?
>
> David VanderVelde
>
> Manager, Liquids NMR Facility, Caltech
>
> davidv_at_caltech.edu
>
>

--
Dr. Bob Berno
Manager, NMR Facility
McMaster University, ABB-156
1280 Main St. West
Hamilton, ON L8S 4M1
T: 905-525-9140 x24158 or x24686
-------------------------------------------------------------------------
This email contains CONFIDENTIAL information intended for the specific recipient(s).
If you have an accommodation need for a planned meeting, please e-mail me directly and
I will do my best to make appropriate arrangements.
Should you require any materials sent via this e-mail address in an alternate/accessible
format, please let me know.

McMaster University sits on the traditional Territories of the Mississauga and
Haudenosaunee Nations, and within the lands protected by the "Dish With One Spoon"
wampum agreement.



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1504): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1504__;!!PvDODwlR4mBZyAb0!U2aDZoB9sOfqMW0JEbRF7SDvlxbGCJ=
d9RlDUoxrJMeFw4Fqjoe5HZPEp0ZTN3x_qFpC6kWFWxE8sjxRP$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/107904283=
/7559972__;!!PvDODwlR4mBZyAb0!U2aDZoB9sOfqMW0JEbRF7SDvlxbGCJd9RlDUoxrJMeFw4=
Fqjoe5HZPEp0ZTN3x_qFpC6kWFWxLDeIhao$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Thu Aug 15 2024 - 05:57:39 MST

This archive was generated by hypermail 2.4.0 : Sun Sep 01 2024 - 15:22:33 MST