Re: [AMMRL] ICON-NMR logins in Topspin 4.4.0

From: Sukenick, George <sukenicg_at_mskcc.org>
Date: Wed, 14 Aug 2024 23:01:01 +0000

I’ve seen something like this (TS 3.6.x; CentOS 7) where there is an “unseen” dialog box -
for some reason, the dialog box, on occasion, is squished down to about the size (maybe smaller?)
than a check box, hard to notice and then it seems like nothing responds because the software
is expecting a response to the totally squished box.
Of course, the solution is to expand the box and dismiss it or answer what it is asking for.

  -------- -------- --------
    George Sukenick, Ph.D.
       Associate Lab Member/Head
       NMR Analytical Core Facility
       Memorial Sloan Kettering Cancer Center


> From: main_at_ammrl.groups.io on behalf of Vander Velde, David
> Date: Wednesday, August 14, 2024 at 6:04 PM
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: [EXTERNAL] [AMMRL] ICON-NMR logins in Topspin 4.4.0

We've been having an issue where ICON-NMR will, seemingly at random, start
nominally accepting logins, but the control buttons remain grayed out, and
the box that says "Automation Locked" does not go away as it usually does
with a login.

We don't see this with TS 3.7.0, on another instrument with virtually the same
user group. I have not been able to associate this with any particular user,
or what they do, or don't do; or whether or not they have an ICON password.

The one thing I'm sure is not involved is the operating system—I saw it first
under CentOS 7, still see the same thing with Alma 9 on a new workstation.

Is anyone else seeing this behavior? If so, has anyone found a workaround other
than restarting the software?

David VanderVelde

Manager, Liquids NMR Facility, Caltech

davidv_at_caltech.edu
_._,_._,_

===================

Please note that this e-mail and any files transmitted from
Memorial Sloan Kettering Cancer Center may be privileged, confidential,
and protected from disclosure under applicable law. If the reader of
this message is not the intended recipient, or an employee or agent
responsible for delivering this message to the intended recipient,
you are hereby notified that any reading, dissemination, distribution,
copying, or other use of this communication or any of its attachments
is strictly prohibited. If you have received this communication in
error, please notify the sender immediately by replying to this message
and deleting this message, any attachments, and all copies and backups
from your computer.

Disclaimer ID:MSKCC


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1501): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1501__;!!PvDODwlR4mBZyAb0!XWKEgj_3spq7NJg_Voezt1xdyDjVuP=
SV_yvuK-QY21wod2DmGBZZ-ZgZKkH58tNuO4LLAyFojWUgJW1R$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/107904283=
/7559972__;!!PvDODwlR4mBZyAb0!XWKEgj_3spq7NJg_Voezt1xdyDjVuPSV_yvuK-QY21wod=
2DmGBZZ-ZgZKkH58tNuO4LLAyFojd_C7AiN$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Aug 14 2024 - 16:01:14 MST

This archive was generated by hypermail 2.4.0 : Sun Sep 01 2024 - 15:22:33 MST