Re: [AMMRL] Off axis shims bad after 24 hours on 40-year-old Oxford magnet #Bruker #Hardware

From: Anklin, Clemens via groups.io <clemens.anklin=bruker.com_at_groups.io>
Date: Thu, 15 Aug 2024 00:43:12 +0000

Dear Bert,

there was an incident in the late 80's where a Oxford 500 round bottom magnet
experienced a catstrophic failure during warm up after energization. liquid
nitrogen entered the vacum chamber and the resulting over pressure led to
the bolts being sheared off at the flange and the top of the magnet including
the coil lifting up quite a bit. Plus the mess of mylar in the lab was epic.

At the time the magnets did not have drop plates. As a fix for these magnets
Oxford exchanged the boltsholding the flange together for some that would
allow a small separation of upper and lower part of the dewar. This would
allow the perssure to escape.

Under normal conditions the vacuum and the bolts held everything together.

This fix was only for magnets that were already installed. All new magnets
after that had drop plates.

With best regards

Clemens


--
==========================
==========================
====================
Clemens Anklin Ph.D. Vice President
Bruker BioSpin NMR Applications & Training
15 Fortune Drive NEW Phone: 978-313-5NMR(5667)
Billerica MA 01821 twitter: _at_canklin
web: http://www.bruker.com
e-mail: clemens.anklin_at_bruker.com
==========================
==========================
====================


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1511): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1511__;!!PvDODwlR4mBZyAb0!WjPDpeWtULbo-Q02dV4nB0SdCNb-iz=
qlbsmu9jEafBqSfeIGh1WMNSj3HekoFl1Q3C_xKaqvsixadZMNGsTwkvpXJgQU$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/107818605=
/7559972__;!!PvDODwlR4mBZyAb0!WjPDpeWtULbo-Q02dV4nB0SdCNb-izqlbsmu9jEafBqSf=
eIGh1WMNSj3HekoFl1Q3C_xKaqvsixadZMNGsTwkkXUaGC8$
Mute #bruker:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mut=
ehashtag/bruker__;!!PvDODwlR4mBZyAb0!WjPDpeWtULbo-Q02dV4nB0SdCNb-izqlbsmu9j=
EafBqSfeIGh1WMNSj3HekoFl1Q3C_xKaqvsixadZMNGsTwki1UOp2D$
Mute #hardware:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/m=
utehashtag/hardware__;!!PvDODwlR4mBZyAb0!WjPDpeWtULbo-Q02dV4nB0SdCNb-izqlbs=
mu9jEafBqSfeIGh1WMNSj3HekoFl1Q3C_xKaqvsixadZMNGsTwkrlERlGw$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Thu Aug 15 2024 - 09:40:12 MST

This archive was generated by hypermail 2.4.0 : Sun Sep 01 2024 - 15:22:33 MST