Re: [AMMRL] Bruker SampleCase error

From: Perry Pellechia <pellechia_at_sc.edu>
Date: Wed, 17 Apr 2024 13:56:12 +0000

Ok, more to add. Ken Josh and David might be right. While tripping the sample
safety bar gives a different initial error in Topspin, the error in Icon when
a sample is submitted with the sample changer “red” is the “System not ready."
So I really cannot be 100% sure that what we are seeing is not the result of the
much more sensitive hight check bar.

This is going to take more testing but we have to put pressure on Bruker to
lessen the sensitivity or at least give us an option. It is way too easy
to tap this bar and bring down the automation.

Perry.
--

Perry J. Pellechia
Director of NMR Services
Research Professor
University of South Carolina
Department of Chemistry and Biochemistry


On Apr 16, 2024, at 7:42 PM, Kenneth Sharp-Knott via groups.io wrote:

It's not possible to roll back the firmware update at the root of the issue.
I tried with a Bruker engineers help.


________________________________
> From: main_at_ammrl.groups.io on behalf of Perry Pellec hia via groups.io
> Sent: Tuesday, April 16, 2024 6:32:40 PM
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: Re: [AMMRL] Bruker SampleCase error

T.J.

Thanks for adding a data point to the issue. I am hesitate to go back to an
earlier version since I made a pretty big jump from 3.6.1 to 3.7.0 and I am
not sure how far back I need to go. This is a case where not having the
software bug reports accessible really hurts the users. I miss those days ...

Perry,



On Apr 16, 2024, at 5:13 PM, Thusitha Jayasundera via groups.io wrote:

Hi Perry--I had this exact issue on 9/15/2023, and reached out to the group
at the time. The only difference was that we upgraded from Topspin 4.1.1 to
4.3 at the time. The solution was, unfortunately, to revert back to 4.1.1, and
there have been zero issues since. There were other suggestions from the
group though (which you should be able to find via a search), but if all
else fails, reverting back to the old version of Topspin should work.

Good luck!

TJ.


Thusitha Jayasundera, Ph.D.
Director, Magnetic Resonance Center
Boston College
Merkert 105
2609 Beacon Street
Chestnut Hill MA 02467-3860

Tel: (617) 552-6984 (office)
Fax: (617) 552-2705<tel:617-552-2705>

Website: https://www.bc.edu/content/bc-web/research/sites/vice-provost-for-research/about/research-facilities/nmr-center


--
Perry.
--

Perry J. Pellechia
Director of NMR Services
Research Professor
University of South Carolina
Department of Chemistry and Biochemistry




-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1155): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1155__;!!PvDODwlR4mBZyAb0!VM2pfg6YZsj_0rM-eHrxLTvRVnZdBz=
dwBoPdATei53Ql6g8NPfMfdCW0tBx-2JkuekF7OlhvixBQ12JjXg$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/105565488=
/7559972__;!!PvDODwlR4mBZyAb0!VM2pfg6YZsj_0rM-eHrxLTvRVnZdBzdwBoPdATei53Ql6=
g8NPfMfdCW0tBx-2JkuekF7OlhvixAzfEN3kg$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Apr 17 2024 - 06:56:27 MST

This archive was generated by hypermail 2.4.0 : Wed May 01 2024 - 15:08:17 MST