Re: [AMMRL] Bruker SampleCase error

From: Perry Pellechia <pellechia_at_sc.edu>
Date: Tue, 16 Apr 2024 22:32:40 +0000

T.J.

Thanks for adding a data point to the issue. I am hesitate to go back to an
earlier version since I made a pretty big jump from 3.6.1 to 3.7.0 and I am not
sure how far back I need to go. This is a case where not having the software bug
reports accessible really hurts the users. I miss those days ...

Perry,



On Apr 16, 2024, at 5:13 PM, Thusitha Jayasundera via groups.io wrote:

Hi Perry--I had this exact issue on 9/15/2023, and reached out to the group
at the time. The only difference was that we upgraded from Topspin 4.1.1 to
4.3 at the time. The solution was, unfortunately, to revert back to 4.1.1,
and there have been zero issues since. There were other suggestions from the
group though (which you should be able to find via a search), but if all else
fails, reverting back to the old version of Topspin should work.

Good luck!

TJ.


Thusitha Jayasundera, Ph.D.
Director, Magnetic Resonance Center
Boston College
Merkert 105
2609 Beacon Street
Chestnut Hill MA 02467-3860

Tel: (617) 552-6984 (office)
Fax: (617) 552-2705<tel:617-552-2705>

Website: https://www.bc.edu/content/bc-web/research/sites/vice-provost-for-research/about/research-facilities/nmr-center


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1152): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1152__;!!PvDODwlR4mBZyAb0!SwRy8eWBps1UTMMv6-OsUXEQaSmeCy=
8CNwlPRJGe6pwyoN3e8FTa92lrfpQW7Ik3Y5ppFmE6Pt_zubJX1A$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/105565488=
/7559972__;!!PvDODwlR4mBZyAb0!SwRy8eWBps1UTMMv6-OsUXEQaSmeCy8CNwlPRJGe6pwyo=
N3e8FTa92lrfpQW7Ik3Y5ppFmE6Pt91iwyHSg$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue Apr 16 2024 - 15:32:51 MST

This archive was generated by hypermail 2.4.0 : Wed May 01 2024 - 15:08:17 MST