Re: [AMMRL] Problem with HPLNA XBB31P 2HP: cannot match

From: Paris Ning, PhD via groups.io <pning=isotope.com_at_groups.io>
Date: Fri, 25 Sep 2026 15:11:53 +0000

Hi Christophe,

I’ve encountered similar issues where the slice endured water damage.
It was the condensation on the hose during nitrogen fill that fell on the to
stack and seeped into the slice. Bruker came in to replace the whole slice
but I figured, as Sebastian mentioned above, fixing the burned components
might be cheaper/faster.

Paris

> From: main_at_ammrl.groups.io On Behalf Of Sebastian Kemper via groups.io
> Sent: Friday, September 25, 2026 7:04 AM
> To: main_at_ammrl.groups.io
> Subject: Re: [AMMRL] Problem with HPLNA XBB31P 2HP: cannot match

Hallo Christophe,

zu deinem aktuellen Problem kann ich nichts sagen, aber wir haben mal einen
Vorverstärker zu Bert Heise geschickt, der zu stark reflektiert hat.
Es waren Schutzdioden und ein Kondensator durchgebrannt und er hat ihn wieder
gefixt.

Viele Grüße,
Sebastian


> Von: main_at_ammrl.groups.io Im Auftrag von Fares, Christophe via groups.io
> Gesendet: Freitag, 25. September 2026 11:05
> An: main_at_ammrl.groups.io
> Betreff: [AMMRL] Problem with HPLNA XBB31P 2HP: cannot match

Dear Spinlanders,

Suddenly I am no longer able to tune/match properly on my X-channel
(probehead : BL4 N-P/H DVT MAS for 500WB). I could isolate the problem
to the HPPR, specifically one of my 2 identical slices (HPLNA XBB31P
2HP 500), i.e. keeping the entire routing the same but switching one
slice for the other: one tunes normally, the other one doesn’t
match all the way down (see screenshots). I tried several nuclei.

Does anyone have experience with this? Is it an easy fix? Otherwise, any
idea where I could get a replacement on the used market?

Cheers

Christophe Fares
MPI-Mülheim


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3797): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3797__;!!PvDODwlR4mBZyAb0!VHsO8KEH2LhLtIx6l1TDP16XQQYkNH=
fbiDvHftvm1h70UVtbaI-v5wZBL5Ud7mSSWcOpnp0plb-fO4niKs32KCQhTGJP$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121423824=
/7559972__;!!PvDODwlR4mBZyAb0!VHsO8KEH2LhLtIx6l1TDP16XQQYkNHfbiDvHftvm1h70U=
VtbaI-v5wZBL5Ud7mSSWcOpnp0plb-fO4niKs32KCJPtK4g$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Fri Sep 25 2026 - 10:25:37 MST

This archive was generated by hypermail 2.4.0 : Fri Oct 02 2026 - 02:30:34 MST