could be the internal reference. Do you have a 50 Ohm terminator. If yes
try "wobb ext50" and follow the instructions.
Clemens
On 9/25/2026 5:04 AM, Fares, Christophe via groups.io wrote:
Dear Spinlanders,
Suddenly I am no longer able to tune/match properly on my X-channel
(probehead : BL4 N-P/H DVT MAS for 500WB). I could isolate the problem
to the HPPR, specifically one of my 2 identical slices (HPLNA XBB31P 2HP 500),
i.e. keeping the entire routing the same but switching one slice for
the other: one tunes normally, the other one doesn’t match all
the way down (see screenshots). I tried several nuclei.
Does anyone have experience with this? Is it an easy fix? Otherwise, any
idea where I could get a replacement on the used market?