Re: [AMMRL] Problem with HPLNA XBB31P 2HP: cannot match

From: Anklin, Clemens via groups.io <clemens.anklin=bruker.com_at_groups.io>
Date: Fri, 25 Sep 2026 12:05:59 +0000

Christophe,

could be the internal reference. Do you have a 50 Ohm terminator. If yes
try "wobb ext50" and follow the instructions.

Clemens


On 9/25/2026 5:04 AM, Fares, Christophe via groups.io wrote:

Dear Spinlanders,

Suddenly I am no longer able to tune/match properly on my X-channel
(probehead : BL4 N-P/H DVT MAS for 500WB). I could isolate the problem
to the HPPR, specifically one of my 2 identical slices (HPLNA XBB31P 2HP 500),
i.e. keeping the entire routing the same but switching one slice for
the other: one tunes normally, the other one doesn’t match all
the way down (see screenshots). I tried several nuclei.

Does anyone have experience with this? Is it an easy fix? Otherwise, any
idea where I could get a replacement on the used market?

Cheers

Christophe Fares
MPI-Mülheim


--
==========================
==========================
====================
Clemens Anklin Ph.D. Vice President
Bruker BioSpin NMR Applications & Training
15 Fortune Drive NEW Phone: 978-313-5NMR(5667)
Billerica MA 01821 bluesky: _at_canklin.bsky.social
web: http://www.bruker.com
e-mail: clemens.anklin_at_bruker.com
==========================
==========================
====================



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3799): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3799__;!!PvDODwlR4mBZyAb0!QudDe7C8K05PfdbYrg6UZw-mZfl76R=
jeVaJGmH3IHgQ0vXD4mjKPa4ge-rhXDW9Ss8yvmlUKq5w4rGQK2T2Vxu84gvno$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121423824=
/7559972__;!!PvDODwlR4mBZyAb0!QudDe7C8K05PfdbYrg6UZw-mZfl76RjeVaJGmH3IHgQ0v=
XD4mjKPa4ge-rhXDW9Ss8yvmlUKq5w4rGQK2T2VxoDU8h0F$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Fri Sep 25 2026 - 13:06:37 MST

This archive was generated by hypermail 2.4.0 : Fri Oct 02 2026 - 02:30:34 MST