Hi Christophe,
Unfortunately that points to the 50 ohm reference in the preamp having failed.
I don't believe this is a repair that can be done in the field. Officially, this
will need to be exchanged--I know they're not cheap. Unofficially, good luck finding
a used one. At least it looks like this is an AVIIIHD console so swapping the
preamp won't require you to cortab the system.
If you've got a 50 ohm terminator and a couple of adapters you can use the command
"atma ext50" to tune the probe if you do need to use both BB preamps. Run the
command and it will prompt you to connect the reference. Disconnect the appropriate
RF cable from the probe and connect the terminator to the end. Then acknowledge
the message in topspin. It will take its reference data and then prompt you to
reconnect the probe. Do this and it will pop up the normal tuning curve and will
either tune if you've run atma, or let you do what you need to if you've used atmm
or wobb.
Cheers,Mike
On Friday, September 25, 2026 at 03:48:37 AM MDT, Fares, Christophe viagroups.io wrote:
Dear Spinlanders,
Suddenly I am no longer able to tune/match properly on my X-channel
(probehead : BL4 N-P/H DVT MAS for 500WB). I could isolate the problem
to the HPPR, specifically one of my 2 identical slices (HPLNA XBB31P
2HP 500), i.e. keeping the entire routing the same but switching one
slice for the other: one tunes normally, the other one doesn’t
match all the way down (see screenshots). I tried several nuclei.
Does anyone have experience with this? Is it an easy fix? Otherwise, any idea
where I could get a replacement on the used market?
Cheers
Christophe Fares
MPI-Mülheim
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3798): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3798__;!!PvDODwlR4mBZyAb0!XmwCNSodoYpkAyRqkQMx_P5krU0jZy=
PzUzdzGnWf3JEoWy2T7Rvsc0jerO0AUUhMsvJQERAE2y8BJfvQHa1prpK-JJni$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121423824=
/7559972__;!!PvDODwlR4mBZyAb0!XmwCNSodoYpkAyRqkQMx_P5krU0jZyPzUzdzGnWf3JEoW=
y2T7Rvsc0jerO0AUUhMsvJQERAE2y8BJfvQHa1prjgVdiTG$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Fri Sep 25 2026 - 10:26:29 MST