I have found a temporary solution for queuing on our new NMR that is close
to what you ask. We are still waiting a visit to finish setting up our
iconnmr.
I am assuming you have activated queuing on your system. The problem is
that the sx command is executed immediately when entered so you have to
force it to the queue with 'qu sx X ' then you need to do edc and setup
your experiment and solvent.
After that you can run within that exp a macro to do
atma
lock -acqu
topshim
rga
zg
I also add a teready 0.1 30 at the start to stabilize temp and o1calib,
pulsecal and special topshim as needed. However for routine experiments you
may not need them
It is not exactly foolproof or very user friendly but it works.
Hope this helps
On Tue, 1 Sept 2026, 22:33 Atterberry, Benjamin via groups.io, wrote:
> Hi all,
>
> I am the NMR Manager at Mississippi State University. Do you by any chance
> have an AU program for Bruker TopSpin that runs samples one after the
> other, like IconNMR does? What I'm looking for is something where you can
> create different datasets in TopSpin, then queue up sample insertion, lock,
> automatic tuning, topshim, getprosol, optimizing receiver gain, and zg.
>
> We have one spectrometer that gets crowded because it has IconNMR, so the
> users like it because they can queue up several 1H experiments. If we could
> use an AU program to "impersonate" IconNMR on another spectrometer, it
> would free up this one instrument.
>
> Best,
> Benjamin Atterberry
>
> ----------------------------------
> *Benjamin A. Atterberry*
> NMR Facility Manager
> Department of Chemistry
> Office: Hand Lab G052
> Mississippi State University
>
>
>
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3734): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3734__;!!PvDODwlR4mBZyAb0!WOQiuN2vF2FvqSY0e17grwcIS1gDMr=
lnlbRWXLE28kJrLacVXVgX1nZaG1vO4oUIdwRdIkWtFZ28sm_n5Fb7q7LDhCZ6$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121039594=
/7559972__;!!PvDODwlR4mBZyAb0!WOQiuN2vF2FvqSY0e17grwcIS1gDMrlnlbRWXLE28kJrL=
acVXVgX1nZaG1vO4oUIdwRdIkWtFZ28sm_n5Fb7q_zUlREH$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Tue Sep 01 2026 - 13:37:52 MST