Re: [AMMRL] AU Program for Automated Sample Queuing in TopSpin?

From: Anklin, Clemens via groups.io <clemens.anklin=bruker.com_at_groups.io>
Date: Tue, 1 Sep 2026 19:56:28 +0000

Hi Benjamin,

why don't you run ICONNMR on the other spectrometer as well? If it is a
question of special experiment not available in the standard collection
in ICON you can always add experiments or you can use the command "sti"
(send to icon) With this you set up the experiment manually then type
sti and define the holder where the sample is. This then submits the
experiment as is to the queue.

Clemens


On 9/1/2026 3:24 PM, Atterberry, Benjamin via groups.io wrote:

Hi all,

I am the NMR Manager at Mississippi State University. Do you by any chance
have an AU program for Bruker TopSpin that runs samples one after the other,
like IconNMR does? What I'm looking for is something where you can create
different datasets in TopSpin, then queue up sample insertion, lock,
automatic tuning, topshim, getprosol, optimizing receiver gain, and zg.

We have one spectrometer that gets crowded because it has IconNMR, so the users
like it because they can queue up several 1H experiments. If we could use
an AU program to "impersonate" IconNMR on another spectrometer, it would
free up this one instrument.

Best,
Benjamin Atterberry

----------------------------------
Benjamin A. Atterberry
NMR Facility Manager
Department of Chemistry
Office: Hand Lab G052
Mississippi State University

Google Scholar: Click here <https://scholar.google.com/citations?user=cn1ylpgAAAAJ&hl=en>


--
==========================
==========================
====================
Clemens Anklin Ph.D. Vice President
Bruker BioSpin NMR Applications & Training
15 Fortune Drive NEW Phone: 978-313-5NMR(5667)
Billerica MA 01821 bluesky: _at_canklin.bsky.social
web: http://www.bruker.com
e-mail: clemens.anklin_at_bruker.com
==========================
==========================
====================



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3733): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3733__;!!PvDODwlR4mBZyAb0!T0C-Ef40eJlJm-wMhOfi9Bl7u44YFJ=
1gDisL9YBBwiYCvAE3oJ3V5L7gzw8s6s5ZOChJaa0tImfFtWvlMuMBf1x7hlN_$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121039594=
/7559972__;!!PvDODwlR4mBZyAb0!T0C-Ef40eJlJm-wMhOfi9Bl7u44YFJ1gDisL9YBBwiYCv=
AE3oJ3V5L7gzw8s6s5ZOChJaa0tImfFtWvlMuMBfz6UtOQ5$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue Sep 01 2026 - 12:58:58 MST

This archive was generated by hypermail 2.4.0 : Fri Oct 02 2026 - 02:30:34 MST